119
7.2
Replacing the Fuse Protecting Powered Relays
CAUTION: Disconnect power to the controller before opening front panel!
Locate the fuse on the circuit board at the back of the controller enclosure under the plastic safety cover. Gently
remove the old fuse from its retaining clip and discard. Press the new fuse into the clip, secure the front panel of
the controller and return power to the unit.
Warning: Use of non-approved fuses can affect product safety approvals. Specifications are shown below. To
insure product safety certifications are maintained, it is recommended that a
Pulse
fuse be used.
Fuse
5 x 20 mm, 6A, 250V
Pulse
P/N
102834
7.3 Replacing the System Fuse (For model codes with relay option 8 or 9 only)
CAUTION: Disconnect power to the controller before opening front panel!
Locate the system fuse on the lower right-hand corner of the circuit board at the back of the controller enclosure
under the plastic safety cover. Gently remove the old fuse from its retaining clip and discard. Press the new fuse
into the clip, secure the front panel of the controller and return power to the unit.
Warning: Use of non-approved fuses can affect product safety approvals. Specifications are shown below. To
en
-
sure product safety certifications are maintained, it is recommended that a
Pulse
fuse be used.
For Relay Option 8: FUSE, 15A, 5x20mm, Fast Acting 250V
Pulse
P/N 104442
For Relay Option 9: FUSE, 20A, 5x20mm, Slow Blow 250V
Pulse
P/N 104443
8.0 TROUBLESHOOTING
CAUTION: Disconnect power to the controller before opening front panel!
Troubleshooting and repair of a malfunctioning controller should only be attempted by qualified personnel using
caution to ensure safety and limit unnecessary further damage. Contact the factory.
8.1 Calibration Failure
Calibrations will fail if the adjustments to the reading are outside of the normal range for a properly functioning sys-
tem. Refer to the instruction manual for the specific sensor being used for further information.
8.1.1 Contacting Conductivity Sensors
The calibration will fail if the adjustment to the gain is outside of 0.5 to 2.0.
Possible Cause
Corrective Action
Dirty electrode
Clean electrode
Improper wiring of sensor to controller
Correct wiring
Wrong cell constant entered
Program the controller cell constant setting at the value that
matches the electrode being used
Incorrect temperature reading or setting
Ensure that the temperature is accurate