DESCRIPTION

 

The dBi HART transducer range has been specified and designed 
to meet the demanding requirements of tod

ay’s

 process level 

measurement applications for liquids and solids. 
All dBi transducers are 2 wire and can either be used in digital 
HART mode or as 4-20mA loop powered devices. The dBi 
transducers are set up using a hart modem with either proprietary 
HART software such as Pactware or Pulsar dBi HART PC software. 
The dBi unit is based on a PZT ceramic transducer element. The 
nominal beam angle is 10 

@ -3 dB (depending on unit). When 

coupled with the DATEM

®

 signal processing they provide 

unmatched performance in industrial process level measurement. 
All dBi transducers include integral temperature compensation.  
Optional submersion shield is available to prevent spurious signal if 
the transducer becomes submerged. 
A range of flange mounting options (ANSI & EN1092) with or 
without a PTFE facing to give improved chemical resistance is 
available. 
An aiming kit is recommended for solids level measurement to help 
focus on the material surface with the angle of repose. 
Standard cable lengths: 

5, 10, 20 or 30m. 

Process Connection: 

1” B

SP, M20 adapter option. 

Operating Temperature: 

-40 to +80°C 

Ingress Protection:   

IP68 

Enclosure material:                  Valox 357 or PVDF. 

 

HAZARDOUS AREA INSTALLATION

 

 

Colour 

Description 

Limits 

RED 

DC Power +ve 

28V DC max. 

BLACK 

DC 0V 

 

GREEN 

Cable Screen 

 

 
 
 
 

 

GENERAL INSTALLATION

 

The dB transducer should be installed directly above the liquid or 
solid level with the transducer axis perpendicular to the surface to be 
measured. 
The transducer can be installed 

using the 1” BSP threa

d on the top of 

the transducer or with the supplied 1” BSP to M

20 thread adapter. 

See figure 1 for examples. 
In some applications, it may not be possible to install the transducer 

using either a flange or the 1” BSP thread, in these circumstances it 

may be possible to suspend the transducer from its cable. In these 
installations, it is recommended that the transducer be secured using 
a chain fitted to one of the chain holes on the top of the transducer, 
see figure 2. 
When installing the transducer avoid aiming the transducer directly at 
fixed obstructions as they may mask the required return echo from 
the liquid or solid level being monitored, see figure 3. 

 

NOTE: Extension cables must use 2 or 3 core screened conductors.  
Capacitance must not exceed 100nF between conductor to screen.  
Resistance should not exceed 40

 per conductor

 

Ex ia

 (Intrinsically Safe) dBi transducers are certified for 

use in hazardous zones 0, 1, 2 (gas) & 20, 21, 22 (dust). 
A certified safety barrier is required. 

 

 

 

Ex mb

 (encapsulated) dBi transducers are certified for use 

in hazardous zones 1, 2 (gas) & 21, 22 (dust). 
No barriers are required but they must be supplied from 
apparatus that provides protection from prospective short-
circuit currents of at least 1500A. 

 

 

 

 

 

HAZARDOUS AREA INSTALLATION, Exia

 

HAZARDOUS AREA INSTALLATION, Exmb

 

X Limitations on use: 

1.  The transducers must be routinely inspected to avoid the build-
up of a dust layer when installed in Zones 20, 21 & 22. 
2.  Electrostatic hazard 

 The transducers must only be wiped with 

a damp or anti-static cloth. 
3.  Only fuses listed on drawing D-804-0978 may be used with the 
Ex mb approved dBi transducers. 

 

 

 

 

 

  Figure 2             Figure 3 

 
          Figure 1 

dBi FLANGED VERSIONS

 

  

 

EN1092 (BS 4504)          

 All dimensions in mm 

Size 

No. holes 

50 

165 

19 

125 

4 -18mm 

80 

200 

19 

160 

8 -18mm 

100 

220 

19 

180 

8 -18mm 

150 

286 

19 

240 

8 -23mm 

200 

337 

19 

295 

12 -23mm 

ANSI Class 150 

Size 

No. holes 

165 

19 

121 

4 -18mm 

200 

19 

152 

4 -18mm 

220 

19 

190 

8 -18mm 

286 

19 

241 

8 -22mm 

343 

19 

298 

8 -22mm 

Flanged versions of the transducer should be installed using plastic bolts with an 
insulating gasket between the transducer flange and the process vessel flange. The 
bolts should not be fully tightened as this may cause acoustic resonance that results in 
increased ring down. A typical flange application is shown above. 

 

Wiring detail 