EN
100~250 bpm
F.
Measurement Performance in Weak Filling Condition
: SpO2 and pulse
rate can be shown correctly when pulse-filling ratio is 0.4%. SpO2 error
is ±4%, pulse rate error is ±2 bpm or ±2% (select larger).
G.
Resistance to surrounding light:
The deviation between the value
measured in the condition of man-made light or indoor natural light
and that of darkroom is less than ±1%.
H.
It is equipped with a function switch
. The product will enter standby
mode when no signal is in the product within 5 seconds.
I.
Optical Sensor
Red light (wavelength is 660nm, 6.65mW)
Infrared (wavelength is 880nm,6.75mW)
5. Accessories
Hanging rope, batteries (optional), user manual
6. Installation
6.1 View of the front panel