11. Function Specification
Information
Display Mode
The Pulse Oxygen Saturation (
SpO
2
)
LCD
Pulse Rate ( PR )
LCD
Pulse Intensity (bar-graph)
LCD
Pulse wave
LCD
SpO
2
Parameter Specification
Measuring range
0%~100%,
(the resolution is 1%).
Accuracy
70%~100%:±2% ,Below 70%
unspecified.
Optical Sensor
Red light (wavelength is 660nm)
Infrared (wavelength is 880nm)
Pulse Parameter Specification
Measuring range
30bpm~250bpm, (the resolution
is 1bpm)
Accuracy
±2bpm or±2% (select larger)
Pulse Intensity
Range
Continuous bar-graph display,
the higher display indicates the