6. Connections
6.1. Cable Connections
•
All cable connections are made to the rear of the instrument
using screwless terminals.
•
Wires must be stripped of insulation by at least 10mm.
•
To install, depress the orange lever beside the terminal
required and push wire into the hole. Release the lever and
pull gently on the wire to ensure it is securely trapped in the
terminal.
6.2. DC Power (DC PWR + , DC PWR –)
•
The DC supply need not be regulated, provided that it is free
of excessive electrical noise and sudden transients.
•
The instrument can be operated from a high quality plug-
pack as long as there is sufficient capacity to drive both it
and the load cells.
6.3. Load Cell Connection
The instrument may be connected for either 4-wire or 6-wire
operation. For more information, refer to
BUILD:CABLE
setting page 17.