Configuration for Set Reference

Remote Loopback or Remote Unit Configuration

Setting Reference

Limit and Connectors Setting

WireXpert supports all configurations specified in TIA and

ISO standards for Fiber certification.  You can configure Set Reference 

from  the SETUP menu and under Test Settings tab. Select Single-Direction

or Bi-Directional setting for the autotest. Followed by selection of 

loopback or remote unit, and the type of reference(one-jumper,

two-jumper, three-jumper). For details of one-jumper, two-jumper 

and three-jumper methods, please refer to users manual. 

Single Directional and Bi-Directional Testing

Local Unit

Remote Unit

Tx

Rx

Tx

Rx

Single Directional Test

Fiber1

Local to Remote Loss

Local to Remote Loss

Fiber2

Remote to Local Loss

Fiber1

Fiber2

Local Unit

Remote Unit

Tx

Rx

Tx

Rx

Bi-Directional Test

Remote Loopback

Remote Unit

Local to Remote Loss

Remote to Local Loss

Local to Remote Loss

Remote to Local Loss

Choose Single Direction unless you have a specific need to perform 

bi-directional testing. In a single Directional measurement, the loss in 

each fiber is measured in one direction only. As shown in the figure 

above, the loss is measured only from Local to Remote on Fiber1 

and Remote to Local direction in Fiber2.

In the case of a Bi-Directional measurement, the loss in each Fiber is

measured in both directions from Local and Remote units. This test 

requires you to swap the fiber connections in the middle of autotest. 

Please don’t swap connections for Setting Reference, just continue both

the steps retaining the connections.

Before performing autotest, the appropriate limits need to be selected for

testing. If either TIA or ISO limits are selected the PASS/FAIL criteria is 

determined by the loss limit specified in the standard. In the case of TIA it is

TIA 568 C.3 and if it is ISO, it is ISO 14763-3 standard.

Link Validation limit allows you to set end-end loss irrespective of the 

configuration such as number of connectors or splices. The test result 

 is determined by the customized loss value configured.

Configure the number of splices and connectors in the link under test.

In the case of a TIA/ISO limit selection, the limit is calculated based on the 

configured number of connectors  and splices. Please exclude the 

connectors in Set Reference.The maximum allowable loss for a connector

is 0.75 dB according to both the standards. Nevertheless,WireXpert allows 

you to set more stringent standards for connector loss should you require. 

This configuration allows selection between loopback testing  or using 

remote unit to perform autotest. In the case of loopback setting, connect  a

short patch cord at the remote end to provide the loopback.

After setting up the configuring  set reference, perform setting reference

from the TOOLS menu and clicking  on SetReference button. 

Use a Reference cord shipped with the kit to perform Reference setting.  

The length and loss of the reference cord determine the quality of reference

set. 

Local Unit

Tx

Rx

Local Unit

Remote Unit

Tx

Rx

Tx

Rx

Note 1: WireXpert prompts an error during autotest if the reference setting

for the corresponding configuration is not performed before running the

autotest.

Note 2: WireXpert also verifies the quality of reference setting and will warn you 

and not allow to complete reference setting if  it detects high loss or long patch

cords. Always use the reference cords shipped with the kit. 