2.10.4 Alarm silence
When alarm generates, press the
【
】
membrane button, the alarm tone will be
suspended, meanwhile, the alarm silence icon
and 100 seconds countdown time
will display in the status information area.
The alarm silence status lasts for 100 seconds, 100 seconds later, the alarm silence will
be cancelled.
Press the
【
】
membrane button again before the 100 seconds period will cancel
the “silence”.
Cautions:
In alarm silence status, except for the alarm tone, the other alarm functions are
working normally.
In alarm silence status, if new alarms occur, the alarm silence will be cancelled and
alarm tone will be activated according to the latest alarm.
Pressing the
【
】
membrane button is invalid if no alarm occurs.
2.10.5 Alarm reset
When alarm generates, press the
【
】
membrane button, all the alarms will be
cleared , the system restarts monitoring.
Pressing the
【
】
membrane button is invalid if no alarm occurs.