background image

 

3.

 

APP Operations

 

 

 

 

Tablet Version 

Mobile Version

 

 

3.1

 

Ultrasound Scanning

 

 

Once the probe is connected, start the app and the main screen will be displayed. No image is 
displayed when the app is opened for the first time. 

 

The WiFi connection status indicates the SSID of the probe connected. Press the FREEZE/LIVE 
button to unfreeze the image and start the diagnosis. Once the image is moving, the GAIN (GN) 
setting can be modified up and down. The top right of the screen displays the date and time, the 
image depth and the Gain in the case of the Smartvus app for tablet, and the top left in the case 
of the SmartVus app for mobile. "LIVE" can be read at the bottom left of the screen when the 
probe is moving, or "FREEZE" is displayed when the image is frozen. 

 

The scan depth (D) can be adjusted by swiping up or down in the middle of the screen. The value, 
which ranges between 100 and180 mm, is displayed below the Gain. There is a green bar below 
that corresponds to the cine loop, and a number to the right that indicates the current image 
recorded. Users can swipe the slider to select the required image. 

 

In the mobile version, 20/20 or100/100 can be stored, depending on your version of the app. To 
switch from one image to another, swipe left or right in the middle of the screen. 

 

The WiFi connection status indicates the WiFi SSID of the probe connected. Click on the animal 
information for a pop-up window to be displayed. Enter the information required in this window: 

 

 

Summary of Contents for SV-3

Page 1: ...Provetscan SV 3 Sector ultrasound scanning for reproductive disgnostic User Manual ...

Page 2: ... is wireless and has no cable The SV 2 probe has a built in ultrasound image processor the power supply and the wireless signal module to connect to the main unit The main unit has now been replaced with a mobile device The probe acts as a WiFi access point and displays the image via the SmartVus app This manual provides an overview and must be read carefully before the device is used 2 ...

Page 3: ...stics Screen iOS or Android mobile device Scanning 80 sector based Probe 3 5 MHz transducer Depth 100 180 mm Display Mode Mode B Scales of grey 256 levels Battery life 5 hours Size 140 mm x 45 mm x 40 mm Weight 200 g ...

Page 4: ...items included Artículos Incluidos SV 3 ultrasound probe USB charging Table Wrist strap User manual Dust proof case Subjection Harness Each item must be examined to notify of any defects or damage that may have occurred during delivery despite it being carefully packed Where this is the case please contact your distributor immediately to inform of the problem 2 2 2 2 Installing the app If the Smar...

Page 5: ... for three seconds to switch the probe off When the probe is switched off the indicator lights will also be off 2 4 Wi Fi Connection When the probe is waiting for a WiFi connection go to settings on the mobile device and search for the probe SSID The SSID is for exampIe SV 3 GMBFLAOO1 Connect the SSID using the password which is the same as the serial number printed on the bottom of the probe The ...

Page 6: ...IVE can be read at the bottom left of the screen when the probe is moving or FREEZE is displayed when the image is frozen The scan depth D can be adjusted by swiping up or down in the middle of the screen The value which ranges between 100 and180 mm is displayed below the Gain There is a green bar below that corresponds to the cine loop and a number to the right that indicates the current image re...

Page 7: ...d 10 seconds on the device Not available on the mobile version 3 4 2 4 Data storage Press the Save data button to save the pregnancy data in the app s memory This data can be displayed by pressing the View data button and can be exported by pressing the ExportFile button to a CSV file so that it can be opened for example with Excel or Numbers The data is deleted by clicking RemoveAll Data can be r...

Page 8: ...op up window will be displayed where the name of the CSV file is entered and is then included in a folder under today s date Several files can be saved on the same day When you open the GSV file you are given the option of opening and sharing the file in other apps such as Excel or Numbers Provided the data is saved it can be edited to modify the diagnosis add comments or delete the record ...

Page 9: ...vent the tablet from becoming dirty with dust and becoming wet or damp a waterproof case is included that is splash and dust resistant We specifically suggest purchasing a case for iPad mini or a case for Galaxy Tab A which also withstands falls of 1 2 m 4 3 Cleaning The probe must be regularly cleaned Use a damp cloth or paper towel The probe has an IPX5 protection rating against water and dust i...

Page 10: ......