- 10 -
5.2 Head-mode
In head-mode the temperature measurement range is 0.0˚C - 100.0˚C or 32˚F - 212˚F.
After any measurement between 32˚C and < 37.6˚C or 89.6˚F and < 99.7˚F one beep
will sound and the display will light green. When the measurement is between 37.6˚C
and < 42.2˚ or 99.7˚F and <108˚F, the fever alarm will sound; 7 beeps will sound and
the display will light red. The measured temperature is shown on the display.
5.3 Out-of-range temperatures
Temperature measurements that are outside the measurement ranges will result in
error codes L˚ or H˚ on the display, accompanied by three beeps and the display
lighting red for three seconds.
“ L˚ ”: Measurement is below 32.0˚C or 89.6˚F in Head-mode or below 0.0˚C or 32˚F
in Ear-mode/ Ambient temperature mode. Please make sure you are following the
instructions on how to get accurate measurements.
“ H˚ “: Measurement is exceeding 42.2˚C or 108˚F in Head-mode or over 100.0˚C or
212˚F in Ear-mode / Ambient temperature mode. Please make sure you are following
the instructions on how to get accurate measurements.
6. Cleaning
•
Regularly clean the thermometer with a damp cloth.
•
Do not use alcohol or benzine for cleaning.
•
Prevent the sensor from getting stained by cleaning the ears before using the
Ear-mode.
•
Prevent the sensor from getting stained by cleaning and drying before using the
Head-mode.
•
Caution: This thermometer is not waterproof! Do not expose this thermometer
to water or any other liquid.
7. Storage
•
Store this thermometer at room temperature; the most accurate temperature
measurement is obtained when the thermometer has been exposed to the same
ambient temperature as the user for at least 30 minutes.
•
Always store this thermometer in the enclosed soft case to prevent it from
collecting dust or getting unhygienic.
8. Troubleshooting
Error: Display shows “ ER1” + backlight is red + 3 beeps sound.
Meaning: Ambient temperature dropped below 10˚C/ 50˚F or exceeded 40˚C/ 104˚F.