1.
Connect the banana plugs to the measuring instrument. Select the corresponding
sensitivity on your instrument. Select AC range if a multimeter is used.
2.
On the flexible current probe, select the desired range (30A, 300A or 3000A) and
output (300mV or 1V).
3.
Connect the flexible probe around the conductor.
4.
Make sure the current flowing direction is in consistent with the arrow marked on the
probe coupling. If the flexible current probe is connected in the right orientation, the
correct phase will be displayed on the oscilloscope.
5.
Keep the probe coupling more than 25mm away from the conductor.
WARNING : Always wear appropriate gloves in operation.
WARNING :
Do not select AC + DC range for measurement. Select AC range.
If a RMS multimeter is used, the AC current value can be calculated from the AC voltage
reading.