INSTRUCTION MANUAL
Flexible AC Current Probe
Model Prova AFLEX 3030
Prova Instruments Inc.
Page 1: ...Prova INSTRUCTION MANUAL Flexible AC Current Probe Model Prova AFLEX 3030 Prova Instruments Inc...
Page 2: ...Warning Page 2 2 Panel and parts description Page 3 3 Operating instructions Page 4 4 Electrical specifications Page 10 5 General specifications Page 11 6 Battery replacement Page 12 7 Maintenance an...
Page 3: ...CAT I Equipment for connection to circuits in which measures are taken to limit the transient over voltages to an appropriate low level Over voltage Category II CAT II Energy consuming equipment to b...
Page 4: ...ing a voltage from 30V to 600V unless you are wearing protective clothing and gloves suitable for high voltage work 2 Always inspect and check any damage of the current probe assembly before usage Do...
Page 5: ...EL AND PARTS DESCRIPTION 1 Coupling assembly 2 Flexible loop 3 Banana plugs for output 4 Output range select switch 5 Power on LED 6 Low battery voltage LED 7 Current range select switch 8 Battery cov...
Page 6: ...3 OPERATING INSTRUCTION 4...
Page 7: ...ke sure the current flowing direction is in consistent with the arrow marked on the probe coupling If the flexible current probe is connected in the right orientation the correct phase will be display...
Page 8: ...Example 1 Current range 300A and output 300mV The AC voltage reading is 100mV The actual AC current 300A 300mV 100mV 100A 6...
Page 9: ...Example 2 Current range 3000A and output 300mV The AC voltage reading is100mV The actual AC current 3000A 300mV 100 0mV 1000A 7...
Page 10: ...Example 3 Current range 300A and output 1V The AC voltage reading is 333 3mV The AC actual AC current 300A V 333 3mV 100A 8...
Page 11: ...Example 4 Current range 3000A and output 1V The AC voltage reading is 333 3mV The actual AC current 3000A 1V 333 3mV 1000A 9...
Page 12: ...conductor located at the center of the flexible loop DC offset 20mVDC 300A AC range 100mVDC 30A Ac range 20mVDC 3000A CA Noise 2mACV RMS Bandwith 3dB 10 10KHz Load impedance 1K Batteries 2 AA 1 5V Alk...
Page 13: ...m Cable length from probe to case 1700mm Cable length from case to output 1700mm Batteries Two 1 5V SUM 3 Dimensions case 130mm L x 80mm W x 43 mm H 5 1 L x 3 1 W x 1 7 H Operating temperature 10 to 8...