Propaq CS Directions for Use
59
Alarms &
Aler
ts
The following messages can appear in the NIBP status window.
CALIBRATING
. The NIBP channel is running an internal calibration.
DISABLED, LOW BATT.
See LOW BATTERY, NIBP DISABLED above.
NIBP DISABLED, SERVICE REQUIRED.
See
SERVICE REQUIRED, NIBP DISABLED
above.
RETRY.
Since the Propaq CS monitor did not receive a valid NIBP reading, it will automatically
attempt to take another reading.
The following NIBP status message looks similar to an equipment alert, although it does not indicate a
malfunction and does not cause an alert tone.
NIBP IN PROGRESS, PLEASE WAIT, FILTERING ARTIFACT.
Noise or artifact such as vehicle
motion is causing a delay while measuring NIBP. To remove the message, press any key below the
screen. To cancel the NIBP measurement, press the
Start/Stop NIBP
key at the right of the screen.
Temperature Messages
PROBE NOT DETECTED.
This message occurs when the Propaq CS monitor has successfully
measured temperature and a probe is then disconnected. Reconnect the probe or acknowledge the
equipment alert by pressing any menu key.
PROBE SHORT.
Verify that the probe is properly inserted in the left side panel. If so, replace
probe.
CALIBRATION ERROR, TEMP DISABLED.
This message appears when the Propaq CS monitor
has detected that it cannot accurately measure the temperature. The monitor should be serviced.
Malfunction of the temperature probes may result in inaccurate readings. Confirm suspect readings.
SpO
2
Messages
SpO
2
messages can appear in the equipment alert window or in the SpO
2
numeric window.
NO SENSOR DETECTED.
Indicates an SpO
2
sensor has been disconnected from the monitor
after being plugged in for more than a few seconds.
SEARCH
: During this search time, the SpO
2
channel tries to detect blood pulsing through the
measurement site. After the measurement has been established, the oxygen saturation value is
displayed in the numeric window.
STANDBY
is displayed in the numeric window when the SpO
2
sensor is disconnected from the
patient, an alarm occurs, and you press the
STANDBY
key. STANDBY is also displayed if you first
plug the SpO
2
sensor cable into the monitor connector before attaching the SpO
2
sensor to the
patient.
Summary of Contents for Propaq 242
Page 6: ...6 Protocol Systems Inc...
Page 18: ...18 Protocol Systems Inc...
Page 26: ...26 Protocol Systems Inc...
Page 50: ...50 Protocol Systems Inc...
Page 52: ...52 Protocol Systems Inc...
Page 64: ...64 Protocol Systems Inc...
Page 66: ...66 Protocol Systems Inc...
Page 72: ...72 Protocol Systems Inc...
Page 74: ...74 Protocol Systems Inc...
Page 78: ...78 Protocol Systems Inc...
Page 84: ...84 Protocol Systems Inc...
Page 86: ...86 Protocol Systems Inc...
Page 94: ...94 Protocol Systems Inc...
Page 120: ...120 Protocol Systems Inc...