![Protek A334M User Manual Download Page 45](http://html.mh-extra.com/html/protek/a334m/a334m_user-manual_1619992045.webp)
1 GENERAL OVERVIEW
45
© 2015 GS Instruments, Co., Ltd
Limit Type Description
Figure
Type
Description
MAX
Based on Limit Line, lower part is regarded as pass
and upper part is regarded as failure..
MIN
Based on Limit Line, upper part is regarded as pass
and lower part is regarded as failure..
Single
Point
Single frequency input is requested without any
start frequency or end frequency. Based on the
input frequency, Pass or failure shall be justified by
the range of Low, High, Limit value.
Table 4.14 Limit Type Description
Figure 4.18 Limit Measure Data in Excel File
Instruction to import Excel file
Setting appropriate Limit list value at Limit Measure Window
Input Model name and tester at Setup frame.
Click Excel Setup button to input DATA at Excel File.
Click Measurement button (It will be active upon Excel set up.)
Measure performance. Result of pass / failure will be immediately
showed and would be saved as well as Excel file.
Summary of Contents for A334M
Page 1: ......