RS-232C REMOTE COMMAND
1. General Features
1) Each Mode Setup
2) Reading and writing each datum
2. COMMAND Type
“
A0 XX CR
”
A0: ASCII CODE
XX: When DATA needed
CR: 0DH(CR)
3. ERROR Output and REMOTE Control
If Communication Error, or Command Error occurs,
“
?
”
will be indicated on the bottom right
side of the screen, and
“
R
”
will be indicated when data receive properly.
[Not] Single time COMMAND process takes greater than 40ms subsequently. Do not use the
command as it possible
4. Signal Line and Pin Number
When it connected to PC, operate it using the MINI DIN Cable (Option)
MINI DIN CABLE
PIN NO
SG
5
1
3
7
4
25PIN DSUB
PIN NO
1
2(3)
3(2)
4(7)
5(8)
6(6)
7(5)
20(1)
(9PIN DSUB)
Signal Name
FG
RXD
TXD
CTS
RTS
DSR
SG
DCD
COMPUTER
PIN NO
1
2(3)
3(2)
4(7)
5(8)
6(6)
7(5)
20(1)
Signal Name
FG
RXD
TXD
CTS
RTS
DSR
SG
DCD
※ ( ) is PIN NO of 9PIN DSUB
5. Interface
Transmitting and receiving data from PC are RTS/CTS type
Not using DSR/DCD, or XON/XOFF
6. Specification
Synchronization: ASYNC
STOP BIT: 2 STOP BIT
PARITY: NONE
DATA Length: 8 BIT
Communication Speed: 1200, 2400, 4800, 9600 BPS
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Summary of Contents for 3201
Page 1: ...99 Washington Street Melrose MA 02176 Fax 781 665 0780 TestEquipmentDepot com ...
Page 2: ...ⅰ ⅰ ...
Page 3: ...ⅱ ⅱ ...
Page 4: ...ⅲ ⅲ ...
Page 14: ...IV FUNCTIONAL DESCRIPTION 1 PANEL DESCRIPTION 9 9 ...
Page 59: ...54 54 ...