PRS 500 X
Instructions for use
5066-0-8002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
34 von 40
7.2
X-ray system tube support, floor stand
Focal spot vertical travel (horizontal X-ray beam):
294 – 1882.5 mm
Floor – Focal spot distance (verticale X-ray beam): max. 1882.5 mm
Rotation X-ray tube assembly around horizontal support arm:
± 180°
detents at:
- 90°, 0°, + 90°, 180°
Vertical travel supporting arm:
1588.5 mm
X-ray tube support, floor stand longitudinal travel :
1411 mm
X-ray tube support, floor stand longitudinal travel,
with short floor rail extension:
2078.5 mm
Extension telescopic X-ray arm (optional):
+230mm
7.3
X-ray system image receptor stand
Standard
Column height:
2190 mm
Vertical shift film center:
385 - 1865 mm
The Bucky, Grid entity and the measure chambers are connected to the generator.
Detailed information please find in the enclosed User Manual, Technical Description of the PROVERT.
7.4
Attenuation Equivalent
WARNING!
The radiographic system PRS 500 X can be delivered with different options on
the Grid Entity/Bucky.
The attenuation factor must be determined at the final inspection at the
customer. The variables like X-ray tube, Collimator etc. have influence to the
factor. The attenuation value of the components can be read out of the
accompanying documents of the component. The attenuation value has to be
determined at the technical specifications. If the limits can’t be kept please
inform PROTEC immediately. If additional accessories are use it has a negative
influence to the quality of the X-ray image.
The cover X-ray system image receptor stand is defined as application part.
The aluminium attenuation equivalent of the cover X-ray system image receptor stand is typically 0,4
and < 0,5 Al mm according to EN 60601-1-3. Tested at 100 kV with a first half-value layer thickness
(HVLT) of 3,7 mm Al.
7.4.1
Protection Art and Protection Class
The PRS 500 X is consistent with a protection class 1 device and contains applicable parts Type B
(according to EN 60601-1).Environmental conditions.
7.5
Automatic cutoff dose
7.5.1
Analogue System
The automatic cutoff dose is 2.5 µGy.
7.5.2
Digital System
The automatic cutoff dose depends on the detector.
For RAPIXX systems, see Installation- & User manual of the corresponding RAPIXX system (Chapter 3.2;
3.3)