12
Rev. 2 / 3-18-2019
PRF-922B Miniature Two-Point Probe Set
Device Tray in Feeder Equipment: This measurement con
fi
rms con nuity from the tray to ground
through several elements
• Paint on tray contact surface
• Support
shelf
• Shelf support and control linkage
• Shelf linkage to machine frame support
• Machine frame to ground
Device Transport Cart: This Rtg measurement combines equipment components with the environ-
ments
fl
oor surface, including:
• Cart shelf surface
• Cart
frame
• Cart
wheels
• Floor surface to ground
Assembly Fixture at Worksta on: Rtg measurement from the
fi
xture may include several elements,
depending on how it is installed, such as:
• The
fi
xture’s device holding assembly
• Fixture
frame
• Worksta on surface suppor ng the
fi
xture
• Worksta on surface to the worksta on’s frame
• Worksta on frame to ground
The photo in Figure 8 is an Rtg measurement of a tray on a worksurface mat. In this con
fi
gura on the
resistance measurement includes the tray, the mat, the mat connec ons and wiring to ground. No ce that
the posi ve test lead is separately connected to ground.
VII. Handling & Maintenance
A. PRF-922B Miniature Two-Point Probe Fixture
1. Store the PRF-922B in a clean, dry environment with both BNC and Probe covers installed for
environmental and mechanical protec on.
2. Periodically, remove all spring-loaded test pins. Clean the spring-loaded test probes and Te
fl
on
moun ng disk with a solu on of laboratory grade isopropyl alcohol and a lint-less cloth, or
laboratory quality swab. Allow components to dry thoroughly be fore re-assembling.
a.
Remove each test probe individually by grasping it
fi
rmly then
pulling it straight out of its socket.
b.
Inspect each probe for damage, then clean with the
alcohol
solu on. If a probe is damaged, i.e., bent, does not
compress smoothly, or has deep surface scratches, replace it
with a new probe of the same size and characteris cs.
(Contact Prostat Corpora on, Customer Service for spare
replace
ment
probes.)
c.
Clean and dry the Te
fl
on moun ng disk twice to insure
cleanliness
d.
Carefully re-install the spring-loaded test probes, and fully
re-seat them in their sockets
Figure 11: Remove Spring-Loaded
probe electrode by grasping
fi
rmly
and pulling straight out of its
socket.
Pull straight out
Summary of Contents for PRF-922B
Page 1: ...PRF 922B User Manual MINIATURE TWO POINT PROBE SET ...
Page 2: ......
Page 16: ...16 Rev 2 3 18 2019 PRF 922B Miniature Two Point Probe Set NOTES ...
Page 17: ...17 Rev 2 3 18 2019 PRF 922B Miniature Two Point Probe Set NOTES ...
Page 18: ...18 Rev 2 3 18 2019 PRF 922B Miniature Two Point Probe Set NOTES ...
Page 19: ......