13
5.
Place the PRS-801-W 5 lbs weight onto the Discharge Arm (Figure 13).
Figure 13. Complete setup for Shielded Bag Test.
6.
Set the Range to HIGH on the PDS-510.
7.
Turn the HIGH VOLTAGE to the ON position.
8.
Use the knob to set the voltage to the level established in System Verification Test.
9.
Click the “Stopped” button at the bottom left hand portion of the screen to place the scope in
“acquisition” mode. The wording “Stopped” will change to “Running” (Figure 7).
10.
Initiate a discharge by pressing the red DISCHARGE button on the PDS-510.
11.
Using the mouse, drag the left scope cursor to the left of the discharge waveform and the
right cursor to the point where the waveform decays to zero. Note: If the waveform dips
below the zero line place the cursor at the point where it crosses the zero line form the
second time (Figure 14).