Prostat PBT-531 User Manual Download Page 1

 

 

 

 

 

USER MANUAL 

SHIELDED BAG TEST KIT AND POWER SUPPLY 

PBT-531 

PDS-510 

Summary of Contents for PBT-531

Page 1: ...USER MANUAL SHIELDED BAG TEST KIT AND POWER SUPPLY PBT 531 PDS 510...

Page 2: ......

Page 3: ...mportant Safety Information 2 INTRODUCTION 4 ASSEMBLY INSTRUCTIONS 6 PRE TEST SETUP 7 SYSTEM VERIFICATION TEST 8 Software Setup 8 SHIELDED BAG TESTING 11 PDS 510 POWER SUPPLY 15 CHARGING THE BATTERY 1...

Page 4: ...ymbol indicates earth ground This symbol represents capacitance This symbol indicates battery safety approval Conforms to European Union directives This product contains a Lithium ion battery that suc...

Page 5: ...be made by an authorized technician Do not exert any pressure on the LCD display from the top In rare cases cleaning the PDS 510 case and display with a slightly dampened cloth may be required Should...

Page 6: ...supply that can be used to charge conductive and dissipative materials WARNING As with any electrical device use proper electrical precautions to avoid personnel shock Operating discharge simulators s...

Page 7: ...is a list of the items included with your product Part No Description PBT 531 PDS 510 PDS 510 HBM Discharge Simulator PBF 520 Shielding Bag Test Fixture CT1 Tektronix CT1 Current Probe PicoScope Osci...

Page 8: ...6 02 ASSEMBLY INSTRUCTION FOR THE PBF 520 Assembly Base Assembly Base Extension Discharge Arm Capacitive Arm Bracket 1 2 3 4 5 A B C D E...

Page 9: ...he Base Assembly A of the PBF 520 fixture Figure 1 Connect the 510 054 red lead to the HBM Figure 2 Connect the other end of the 510 054 lead to the Output connector on the PDS 510 Discharge Arm Conne...

Page 10: ...TEST Software Setup 1 Install the PicoScope software onto the computer by downloading it from Pico s website at https www picotech com downloads Use the version that is closest to the model of your P...

Page 11: ...000 volts 10 In the PicoScope software if needed move the yellow trigger icon above the scope s zero line 0 2A on the left vertical scale Figure 7 11 Ensure that the scope horizontal scale is set to...

Page 12: ...lated energy should be 50 6 uJ microjoules and the peak current should be greater than 420 mA milliamps 0 42 amps The blue line is the discharge waveform and the red line is the calculated energy curv...

Page 13: ...Open and select the Shielding Bag Test File or Shielding Bag Test File with Manual Cursors file from the desktop NOTE If you wish the software to analyze the test automatically use the Shielding Bag T...

Page 14: ...p of it lines up with the mark on the Capacitive Arm Figure 11 Figure 10 The Capacitive Arm inside the shielded bag Figure 11 The end mark on the Capacitive Arm represents 10 cm 4 from the top of the...

Page 15: ...utton at the bottom left hand portion of the screen to place the scope in acquisition mode The wording Stopped will change to Running Figure 7 10 Initiate a discharge by pressing the red DISCHARGE but...

Page 16: ...e calculated energy 13 Repeat the process five more times for a total of six discharges per bag 14 Repeat the test for five more bags 15 ANSI ESD S11 31 requires that the discharge energy for static s...

Page 17: ...teries to charge make sure the Main Power switch is ON The batteries will not charge if the Main Power is OFF The charger comes with several adapters for different outlets 08 TROUBLESHOOTING PicoScope...

Page 18: ...or 3 5 or later by going to the Add Remove Programs window in Control Panel and verifying which one is installed 9 Ensure all the latest Windows updates have been applied to the PC 10 Is the PC conne...

Page 19: ......

Page 20: ...e owners Complying with all applicable copyright laws is the responsibility of the user Modification of this document is not permitted without written permission from Prostat Corporation Content subje...

Reviews: