MAINTENANCE
Beyond replacing batteries and fuses, do not
attempt to repair or service your Meter unless you
are qualified to do so and have the relevant
calibration, performance test and service
instruments. The recommended calibration cycle is
12 months.
To clean the terminals
a) Push the Meter OFF and remove test leads.
b) Shake out any dirt that may be in the terminals.
c) Soak a new swab with isopropyl and work around
the inside of each input terminal
d) Use a new swab to apply a light coat of fine
machine oil to the inside of each terminal.
FUSES TEST
Warning
To avoid electric shock or injury, remove the test
leads and any input signals before replacing the
fuses
1. Turn the rotary switch to 200mA position.
2. Use a multimeter to measure resistance of
"V
Ω
mA" terminal or 10A terminal to COM
terminal.
A good mA terminal or 10A terminal fuse is
indicated by a reading between 0
Ω
and 10
Ω
.
13
If the display is overloaded, replace the fuse and
Summary of Contents for MT-1233C
Page 16: ...16 MT 1233C D 3 1 2 1 1 2 1 3 MT 1233C 1 MT 1233C D IEC61010 CAT I 600V CAT II 300V 2...
Page 17: ...1 2 3 500V 4 60V 42V rms 5 6 7 8 10A 9 10 11 12 17...
Page 18: ...13 AC DC AC DC 500Vrms 10A mA F200mA 250V 1999 3 1 H 18 l 5V AAA x2...
Page 19: ...0 40 32 104 10 50 C 14 122 130L 72W 28 H 130 1 2 3 4 5 6 10A 7 1 HOLD 2 10 19...
Page 20: ...1 a V mA COM b c 500V 1 10M 10K 0 1 2 a V mA l0A COM b c mA 5 20mm F200mA 250V 10A 10 15 20...
Page 21: ...3 a V mA COM b c 200 0 5 4 a V mA COM b c d 50 PN 21...
Page 22: ...0 6V 1 5 MT 1233C a V mA COM b c K 250 60V 30V 6 MT 1233D V mA COM 50Hz 1M 3V 10V 22...
Page 26: ...1 OFF 2 3 AAA 1 5 2 4 26...
Page 27: ...27...
Page 28: ...28...