FOS/TAC 2000
TECHNICAL SPECIFICATIONS
812-04.2015-EN
FOS/TAC 2000 User Manual
23
11.1
Technical specifications
FOS/TAC
analyzer
Parameter
Measuring range
Resolution
Measuring method
Other
pH
0 … 10 pH
0.001 pH
Combination electrode
Temperature
0 … 50 °C
0.01 °C
NTC
FOS
1 … 10 g/kg
0.01 g/kg
Titration
Unit: g
org.acids
/ kg
substrate
TAC
5 … 20 g/kg
0.01 g/kg
Titration
Unit: g
CaCO3
/ kg
substrate
FOS/TAC
0.05 …2.00 -
0.01 -
Measured value calculated
Device / analyzer
System casing
Steel sheet casing, powder-coated
Dimensions (w x h x d)
405 x 290 x 160 mm
Weight
Approx. 7.5 kg (including 1000ml of titration solution)
Protection
IP 50 according to DIN 40050
Display
LCD of the IS7430, two-line (pH and temperature)
LCD of the FOS/TAC controller, four-line (FOS, TAC and FOS/TAC)
Operation / input
Keyboard of the FOS/TAC controller
Titration solution / standard
1 litre / measuring solution 0.1N H
2
SO
4
/ sufficient for approx. 30 FOS/TAC measurements
Requirements for the place of installation
Operating temperature
10 to 40°C
Storage temperature
2 to 50°C
Rel. humidity
<=90% annual average, minor and seldom condensation when switched off
Electrical data
Voltage supply, plug-in power supply unit
100-240VAC / 47-63 Hz / 1A @230V / Euro plug / (included in the scope of delivery)
Voltage output
24 VDC /
1 A / 24 W
Connector
Hollow connector 5.5x2.1x11 mm / centre positive
Summary of Contents for FOS/TAC 2000
Page 2: ......
Page 4: ......
Page 21: ...FOS TAC 2000 APPENDIX 812 04 2015 EN FOS TAC 2000 User Manual 21 11 Appendix Appendix ...
Page 22: ......