Pronk Technologies SimSlim SL-8 Operator'S Manual Download Page 10

Detailed Cardiac Output Operation 

•  Connecting the Cables 

The SimSlim simulator is designed to operate without a custom C.O. cable.  Instead,  use the 
monitor's standard C.O. cable.  The connections are as follows: 

1. Connect the monitor's C.O. cable to the monitor. 

2. Connect the catheter connector on the C.O. cable to the C.O. port on the SimSlim simulator. 
3.  Make sure the normal injectate temp sensor is connected to the C.O. cable.  The injectate 
temp sensor should be left at room temperature.  It does not need to be be placed in a bath of 
any kind, but for best results, place it somewhere the temperature will be room temperature and 
relatively stable (i.e. not directly, under a hot lamp or in your hand). 

• Reading the lnjectate Temp 

With  the  cables  connected  correctly,  you  should  be  able  to  read  the  room  temperature  as 
measured by the injectate probe from the monitor. 

• Looking up the Computation Constant (C.C.) Value 

Once you know the current injectate temp, you can look up the C.C. value from the table on the 
back of the SimSlim simulator.  Once you have the correct C.C. value, enter it into the monitor. 

The granularity of the C.C. table on the SimSlim simulator can introduce up to a 5% error in the 
readings.  This should be  fine for most  applications,  but if you  want better precision you can 
interpolate between values.  For example, if your injectate temp was 25.5C, you would look up 
the value for 25 (0.500) and the value for 26 (.545) and split the difference (0.522). 

If you want to use an injectate temp value that is not in the table, you can calculate the correct 

C.C. value according to C.C.

=

6/(37-Ti). 

• Selecting the Simulation mode 

Referring  to the label on the side of the SimSlim,  navigate the modes by double clicking the 
MODE button to select the proper mode to simulate the intended 3, 5 or 7 1/m simulation. 

PRC:,NK  TECHNC:,LC:,C31ES 

Revision 

L

 0

7/18/

2017

 

www.pronktech.com 

(800) 541-9802 

Page 10 

Summary of Contents for SimSlim SL-8

Page 1: ...SimSlim TM Patient Simulator Operator s Manual PReNK TECHNOLOGIES www pronktech com 2014...

Page 2: ...Revision L 07 18 2017 www pronktech com 800 541 9802 Page 2 Compliant products conforming to the relevant standards and directives required for CE have a CE Mark label as shown below affixed to the u...

Page 3: ...eration 6 Detailed Invasive Blood Pressure Operation8 Detailed Cardiac Output Operation 10 Detailed YSI Temp Operation 12 Troubleshooting Tips 13 Services 13 Warranty and Repair Information 14 Specifi...

Page 4: ...Getting Started please refer to each parameter section in this manual for more detailed operational information 1 Turning the SlmSllm Simulator on You never have to The SimSlim simulator runs all day...

Page 5: ...that 15 seconds After 15 seconds the blood pressures will return to their prior values You may shorten the 15 second wait time to resume IBP simulation mode by pressing the MODE button once any time a...

Page 6: ...he snap such that the metal part of the plug is against the underside of the snap Two different size grooves are provided for different size banana plugs ECG Size ECG size is 1mV Lead II in ECG NSR MO...

Page 7: ...1 2 MODE Revison L 07 18 2017...

Page 8: ...t 800 541 9802 for more information The IBP simulation ports are labeled on the bottom of the unit Zeroing Simulated Blood Pressures In all modes except IBP STEPS clicking the mode button will force a...

Page 9: ...c 200mmHg on all four channels IBP STEPS MODE In this mode a selection of static blood pressures are available on all four channels The pressures available are shown on the bottom of the unit and are...

Page 10: ...e from the monitor Looking up the Computation Constant C C Value Once you know the current injectate temp you can look up the C C value from the table on the back of the SimSlim simulator Once you hav...

Page 11: ...to use an injectate temp simulator rather than the actual injectate temp probe Unfortunately different monitor vendors use such a wide variety of thermistors and connectors for injectate temperature t...

Page 12: ...1 4 1 4 monitors cable and a 3 5mm to 1 4 stereo plug converter Revision L 07 18 2017...

Page 13: ...some monitors will occasionally ECG waveform have interaction between the IBP and ECG Unplug IBP to work around the problem On E7 hardware touching the metal shell of an IBP connector and an ECG snap...

Page 14: ...liable only for repair or replacement of the SimSlim Simulator and optional features Pronk Technologies shall not be liable for any incidental or consequential damages Order Cancellation and Refund Po...

Page 15: ...PRc NK TEOHNc Lc C31ES Revision L 07 18 2017 www pronktech com 800 541 9802 Page 15...

Page 16: ...PRc NK TEOHNc Lc C31ES Revision L 07 18 2017 www pronktech com 800 541 9802 Page 16...

Page 17: ......

Page 18: ......

Page 19: ......

Page 20: ......

Reviews: