background image

If this is the case you cannot use this method.

Neo Mode Simulation Setup

    In a neonatal NIBP setup the patient cuff may 

represent  just  a  few  ml  of  volume,  while  the 

patient hose may represent 50 ml or more of 

volume.  Since the SimCube operates by pulse 

volume (about 1/4 ml at peak in neo mode) the 

pressure pulses can be much larger on single 

lumen  than  on  dual  lumen  systems.  Some 

single  lumen  systems  do  not  have  enough 

dynamic  range  to  reliably  capture  the  pulses 

produced by the SimCube with a #1 neo cuff.   

While we have had good results with a #2 neo 

cuff, we recommend using a loosely wrapped 

#3  neo  cuff  for  neonatal  simulation  on  single 

lumen  systems.  Dual  lumen  systems  should 

work reliably with a #1 neo cuff.

  Due to the reduced pulse size of our neonatal 

simulation many monitors will not be able take 

readings on this simulation if the monitor is in 

adult mode or if you are using an adult cuff.  We 

strongly recommend using a neonatal cuff and 

placing  the  monitor  in  neonatal  mode.  

However,  if  you  do  not  have  access  to  a 

neonatal cuff, any volume in the range to 15 to 

75ml will work with most monitors.  This can be 

achieved with an infant cuff or, in some cases, 

with a tightly rolled child cuff.   Also, a length of 

tubing  can  be  used.    A  rough  but  handy 

approximation is that a 1 inch length of 1/8 inch 

I.D. tubing is around 1ml in volume.

Reading to Reading Variations

  Even with the best setup there is still variation 

from reading to reading.  Sometimes a reading 

will be produced which is quite far off target and 

it is difficult to know if this represents a problem 

with  the  monitor  or  not. The AAMI  spec  says 

that  1%  of  readings  can  be  as  much  as 

29mmHg  off,  but  under  simulation,  and  using 

the techniques we have described you should 

see much less than that.  The following chart 

shows some typical error distributions.  These 

represent  the  difference  from  120mmHg 

systolic using SimCube's adult 

mode  with  a  400ml  Cuff,  discarding  the  first 

reading,  for  50  readings.    On  some  monitors 

systolic  and  diastolic  have  similar  error 

distributions,  while  on  others  they  are  quite 

different.

!

"

# !

# "

$ !

$ "

% !

% "

& !

& "

MDE

HP

Critikon

<2mmHg
2-3mmHg
4-5mmHg
6-7mmHg
8-9mmHg
10+mmHg

About The Author

Karl Ruiter has been an engineer, a manager, 

and a director at a Los Angeles-based medical 

device  manufacturer  where  he  designed  a 

NIBP  algorithm  which  is  in  use  in  tens  of 

thousands of units worldwide.  This algorithm 

has been shown by competing manufacturers 

to  be  one  of  the  most  consistent  and 

noise-immune  on  the  market.  In  2003  he 

helped to found Pronk Technologies Inc, which 

bases  its  product  line  on  the  concept  that 

biomedical  engineers  do  important  work  and 

they should have really good tools.   He was 

the lead engineer on the SimCube product.  

More Info

More  information  about  the  SimCube  and 

NIBP simulation can be found:   

On the Web at:

www.pronktech.com

By email at:  

[email protected]

By phone at: 

(800) 609-9802

Revision 

O

 0

7/18/

2017 

 www.pronktech.com  

800-541-9802

Page 20

PRONK TECHNOLOGIES

Summary of Contents for SimCube

Page 1: ...SimCube NIBP Simulator Operator s Manual www pronktech com...

Page 2: ...orming to the relevant standards and directives required for CE have a CE Mark label as shown below affixed to the unit CE Marked products also comply with the WEEE directive for category 9 Monitoring...

Page 3: ...18 2017 www pronktech com 800 541 9802 Page 3 PRONK TECHNOLOGIES Getting Started 4 NIBP Detailed Information 6 ECG RESP Detailed Information 8 IBP Detailed Information SC 5 Only 9 Battery Operation 1...

Page 4: ...e are printed on the front panel next to the LED horizontally For more detailed information on each mode refer to each parameter section in this manual 5 For dynamic NIBP testing connect the SimCube s...

Page 5: ...supplied with the SimCube system and the correct IBP adapter for the monitor under test to the SimCube simulator A variety of IBP adapters are available from Pronk Technologies Inc Select the 0 mmHg...

Page 6: ...scribed below Refer to the front panel of your SimCube simulator to determine which modes apply to your model NIBP Adult 120 80 Simulates a patient with blood pressure of 120 80mmHg heart rate of 70 b...

Page 7: ...ses the monitor must be placed in a special test mode so that it will not release the cuff pressure NIBP Leak Test SC 5 only This mode is used for measuring leak rates in NIBP monitors their cuffs and...

Page 8: ...tion mode SC 5 only The Arrhythmia mode is designed to allow you to quickly check that all the life critical ECG alarms on a patient monitor are properly configured The mode consists of a single cardi...

Page 9: ...he NIBP values associated with the same mode and are indicated by the arrow on the front panel in the IBP column pointing towards the NIBP column IBP 0 Use this mode to simulate atmospheric pressure a...

Page 10: ...g features in your new SC 5 as compared to any previous versions you may have The display will be slightly dimmer The SimCube will go into sleep mode when the NIBP functions are not in use the display...

Page 11: ...that the switch is in the Batt OFF position when you are done the numeric display on the SimCube simulator is blanked under certain conditions to conserve power so you cannot rely on that to indicate...

Page 12: ...SimCube System Troubleshooting Tips Revision O 07 18 2017 www pronktech com 800 541 9802 Page 12 PRONK TECHNOLOGIES...

Page 13: ...noise by running the SimCube on battery power to see if the noise still exists If it is present check to see that the monitor under test has its A C filter s enabled Please note that monitors may hav...

Page 14: ...tors This cable has a six pin mini DIN to GE style connector It is six inches long and works with the IBP Extension cable IBP Interface Cable for Datex Ohmeda Monitors This cable has a six pin mini DI...

Page 15: ...s an annual calibration interval Extended Warranty up to 5 years total may be purchased only at time of original sale or with purchase of Rejuvenation Service Revision O 07 18 2017 www pronktech com 8...

Page 16: ...se or use Pronk Technologies shall be liable only for repair or replacement of the SimCube NIBP Simulator and optional features Pronk Technologies shall not be liable for any incidental or consequenti...

Page 17: ...l readings but 5 might be only within 21mmHg and 1 might be as far out as 29mmHg While these specifications may seem very loose they can be quite difficult to meet The errors in the readings come from...

Page 18: ...at they should be statistically identical but in practice they can still give different results 2 Every manufacturer has adjusted their system based on data which was measured by the nurses they used...

Page 19: ...ormally work with a broad range of test volumes but with smaller test volumes you may see more vendor to vendor reading differences 2 Be prepared to throw out the first reading We find that some vendo...

Page 20: ...h the best setup there is still variation from reading to reading Sometimes a reading will be produced which is quite far off target and it is difficult to know if this represents a problem with the m...

Page 21: ...Revision O 07 18 2017 www pronktech com 800 541 9802 Page 21 PRONK TECHNOLOGIES...

Page 22: ...Revision O 07 18 2017 www pronktech com 800 541 9802 Page 22 PRONK TECHNOLOGIES...

Page 23: ...Revision O 007 18 2017 www pronktech com 800 541 9802 Page 23 PRONK TECHNOLOGIES...

Page 24: ...erent models and sometimes even different software versions can give quite different readings but the following are some average values obtained from a variety of devices Revision O 007 18 2017 www pr...

Reviews: