Indexing the standard solution with your cuvette
CAUTION!
Malfunction of the control circuit
Possible consequence: Poor end-product quality
Ensure that your controller or other measuring equipment is configured so that the
indexing process does not lead to unintentional control functions in the measuring
and control system. The controller connected to the DULCO
®
turb C may not process
the signals issued by the DULCO
®
turb C and use them for control purposes while
the signals are being indexed.
NOTICE!
Screen resolution
For the purposes of indexing, we recommend that the DULCO
®
turb C should be set
in the "Commissioning/solution display" menu so that the following resolutions are
obtained
–
1000 NTU: No digits after the decimal point on the display
–
10 NTU: Two digits after the decimal point on the display
–
0.02 NTU: Four digits after the decimal point on the display
NOTICE!
Freezing of standard solution
The standard solutions must not freeze. Even for short periods of time.