background image

Quick Protocol

Instructions for Use of Product AS1831.

Maxwell

®

 RSC Enviro Total Nucleic Acid Kit 

Wastewater sample (40ml).

1. Add Binding Buffer 1 to each tube.
2. Add Binding Buffer 2 to each tube.
3. Add isopropanol to each tube. Mix.

Add Protease Solution, centrifuge, 

then divide supernatant 

equally into two tubes.

Transfer liquid to Reservoir Extension 

Funnel on PureYield™ Binding Column.

Apply vacuum.

Place microcentrifuge tube in 

Eluator™ Device, add Nuclease-Free 

Water and apply vacuum to elute 

nucleic acid.

1. Add Column Wash 1, apply vacuum.
2. Add Column Wash 2, apply vacuum. 

17691M

A

Standard Protocol for Capture, Concentration and    

 

Clean-Up

Direct Capture and Concentration

1.  Add 40ml of pasteurized wastewater into a 50ml conical tube.
2.  Add 0.5ml of Protease Solution. Mix by inversion and incubate for 30 minutes 

at ambient temperature. 

3.  Clarify sample by centrifuging at 3,000 × 

g

 

for 10 minutes.

4.  Carefully decant 20ml of the supernatant into each of two new 50ml conical 

tubes.

5.  To each tube containing 20ml of the clarified supernatant, add 6ml of Binding 

Buffer 1 (BBD) followed by 0.5ml of Binding Buffer 2 (BBE). 

6.  Mix well by inversion.
7.  Add 24ml of isopropanol to each tube. Mix well by inversion. 
8.  Attach a Reservoir Extension Funnel to the PureYield™ Binding Column, then 

connect the column to the vacuum manifold. 

9.  Pour the mixture from each tube from Step 7 into the Reservoir Extension 

Funnel on the PureYield™ Binding Column. 

10.  Turn on the pump and apply vacuum to capture total nucleic acid on the 

column.

11.  Add 5ml of Column Wash 1 (CWE) and apply a vacuum to pull the liquid 

through the PureYield™ Binding Column.

12.  Add 20ml of Column Wash 2 (RWA) and apply a vacuum to pull the liquid 

through the PureYield™ Binding Column. 

13.  Continue the vacuum for an additional 30 seconds after all liquid has passed 

through the membrane.

14.  Release the vacuum and remove the column from the vacuum manifold.
15.  Assemble the elution device by placing a 1.5ml microcentrifuge tube into the 

base of the Eluator™ Vacuum Elution Device.

16.  Place the Eluator™ Device assembly onto a vacuum manifold. 
17.  Add 250μl of preheated (60°C) Nuclease-Free Water to the PureYield™ Binding 

Column. Apply maximum vacuum until all liquid has passed through the column.

18.  Repeat the elution by adding another 250μl of preheated Nuclease-Free Water to the 

PureYield™ Binding Column.

(continued)

PAGE 1 

 PART #FB237

Figure 1. Schematic showing direct capture of nucleic acid 
from wastewater using a PureYield™ Binding Column.

Reviews: