MAINTENANCE MANUAL //

 STRETCHER PC-750 + RAIL PB-720

Revision 2021/10

5

 

Check the fixation of the bed joints: these joints must be free to be able to carry out the movement, but 

without the assembly losing cohesion. If this joint is not correctly assembled, it could cause damage 

during use, in which case the Promeba technical service should be contacted to consider whether it 

needs to be replaced.

 

Check that the screws that connect the bed to the stretcher are properly tightened.

2.1.7 GENERAL:

 

Check that all screws not specified above are properly tightened.

2.2 RAIL

2.2.1 GENERAL:

 

Check that all the screws are properly tightened.

02 

EVERY YEAR

Summary of Contents for PB-720 Series

Page 1: ...MAINTENANCE MANUAL ELECTRIC STRETCHER SERIES PC 750 RAIL SERIES PB 720 Revision 2021 10...

Page 2: ...e screws on both sides of the lower structure see Fig 1 The tightening torque must be 19 11 Nm Clean the guides on both sides and renew the lubrication with Vaseline see Fig 2 IMPORTANT after greasing...

Page 3: ...ee Fig 4 Check that all the elements screwed on top of the rail are fixed central hooks front lock loading head etc Open the manhole covers clean and grease the lateral move ment guides with Vaseline...

Page 4: ...tor when raising and lowering the stretcher and confirm that it is working well without exerting any extra effort 2 1 5 LOADING HEAD Check the return of the load terminals of the head press each of th...

Page 5: ...nt is not correctly assembled it could cause damage during use in which case the Promeba technical service should be contacted to consider whether it needs to be replaced Check that the screws that co...

Page 6: ...ery two years or according to wear Check the condition of the telescopic lever handles and their inner profiles and consider whether they need to be replaced These elements must be replaced every two...

Page 7: ...r it is necessary to replace them 4 2 RAIL 4 2 1 FIXED RAIL OPERATION Check the condition of the adhesives and consider whether it is necessary to replace them 4 2 2 FIXED RAIL OPERATION Check the con...

Page 8: ...derecho ha modificaciones sin previo aviso Promeba S L no se considera responsabe de los da os causados por la falta o la inexactitud de la informaci n aqu mencionada All rights reserved Variations ca...

Reviews: