3.6  Measuring with Pundit PL-200PE

NOTE! Testing with Pulse Echo Technology requires in-depth 
knowledge of the test object and application characteristics. 
Proceq offers comprehensive ultrasonic training seminars 
imparting this knowledge as well as all functionalities and 
features of the Pundit instruments. Pundit PL-200PE users 
are recommended by Proceq to register for the Advanced 
Ultrasonic Tomography Applications training. Details can be 
found on the Proceq website. 

Preparation

•  Very few preparations are necessary when measuring with the Pulse 

Echo Transducer.

•  The dry-contact transducer means that acoustic coupling is ensured 

without the use of any couplant.

•  The contacts are sprung and can accommodate surface irregularities 

to a depth of 7 mm, so smoothing of the surface is also unnecessary.

•  For B-Scans a test grid should be drawn out on the surface. Alterna-

tively Proceq provide a calibrated tape (Part No. 327 010 71) that can 
be stuck on the surface for the test and then removed afterwards.

Calibration

The most accurate results can be obtained if the instrument is calibrated 
for the material under test.

•  This is done by performing a pulse velocity measurement on a section 

of the structure of known thickness.

•  Alternatively, perform the automatic estimation of pulse velocity as de-

scribed in chapter 3.4.

Physical influences on the Measurement

Inhomogeneities (e. g. aggregate particles, voids) in concrete influence 
the propagation of an ultrasound pulse. They will scatter the signal. The 
effect is very large if the size of the aggregate is equal to or larger than the 
wavelength of the ultrasonic signal. This influence is significantly reduced 
when the wavelength is at least twice as large as the aggregate size. It 
also follows that it is very difficult to detect an anomaly if it is smaller than 
half the wavelength.
The Pulse Echo Transducer is a shear wave transducer with a frequency 
of 50 kHz.
Assuming a typical pulse velocity of 2500 m/s, the wavelength is 50 mm. 
This means that anomalies smaller than 25 mm will be invisible.

Effect of sample size

The geometry of the object is very important for obtaining good results.

•  The maximum penetration depth depends on the quality of the con-

crete and also on the amount of rebars present. Typically the maximum 
transmission depth lies between 50 cm (19.7”) and 1m (39.4”).

•  Generally, the minimum lateral dimension should be two times the 

thickness of the object, or the depth of the anomaly you are trying to 
detect. The reason for this is that if the object is too narrow, reflections 
from the side walls will interfere with the echo from the rear wall.  

4.  Explorer

From the main menu select Explorer to review saved files.

Tap on a saved file to open it.
Return to the Explorer list by press-
ing the back button.
To delete a file tap in the check box 
to the left of the file and delete it

Summary of Contents for Pundit PL-200PE

Page 1: ...Made in Switzerland 60 Years of Innovation Pundit PL 200PE OPERATING INSTRUCTIONS Made in Switzerland 60 Years of Innovation...

Page 2: ...ucer D Pundit Pulse Echo Cable E Pundit Pulse Echo Contact Tester F Battery Charger G USB Cable H DVD with Software I Documentation J Carrying Strap K Calibrated Tape Part of the Pundit Pulse Echo Tra...

Page 3: ...9 9 9 9 9 9 9 9 9 9 23 2 3 9 3 2 23 3 2014 Proceq SA Overview Pundit PL 200PE 1 1 2 3 4 5 6 7 8 9 10 2...

Page 4: ...3 5 Measurement Modes 11 3 6 Measuring with Pundit PL 200PE 12 4 Explorer 12 4 1 Analysis of B Scans 13 5 Ordering Information 13 5 1 Units 13 5 2 Transducers 13 5 3 Accessories 14 6 Maintenance and S...

Page 5: ...ents Serious damage resulting from the effects of foreign bodies accidents vandalism and force majeure All information contained in this documentation is presented in good faith and believed to be cor...

Page 6: ...Transducer Frequency 50 kHz Aperture Size 2x25 cm2 3 Operation The information provided in this manual covers the Pulse Echo application only The Pundit PL 200PE also supports all functionality provi...

Page 7: ...ain menu is displayed All functions may be accessed directly via the touchscreen Return to the previous menu by pressing the back button or the return icon arrow at the top left of the touchscreen Mea...

Page 8: ...a quick read out of slab thickness is desired For complex objects containing internal defects pipes and rebars it is recommended to perform a complete B Scan for analysis of the object Filter Filters...

Page 9: ...g a measurement Zoom out by placing thumb and index finger apart on the screen and pinching them together Pan Pan the image from left to right by dragging Measuring screen controls see page 3 1 Filena...

Page 10: ...ows a marker to be set on a second echo and the difference between thet two echos is displayed Manually adjust the colour intensity 10 Automatic Estimation of Pulse Velocity This setting is available...

Page 11: ...on the transducer The LED s on the transducer provide a visual feedback when a meas urement has taken place successfully This is accompanied by an acoustic tone on the display device The volume of th...

Page 12: ...er 3 4 Physical influences on the Measurement Inhomogeneities e g aggregate particles voids in concrete influence the propagation of an ultrasound pulse They will scatter the signal The effect is very...

Page 13: ...ink once the files have been downloaded to a PC 5 Ordering Information 5 1 Units PART NO DESCRIPTION 327 10 002 Pundit Touchscreen without transducers Consisting of Pundit Touchscreen BNC adapter cabl...

Page 14: ...ectromagnetic radiation produced by the instrument 6 Maintenance and Support 6 1 Maintenance To guarantee consistent reliable and accurate measurements the instru ment should be calibrated on a yearly...

Page 15: ...ct the language and the date and time format to be used Connecting to a Pundit Touchscreen Unit Connect the Pundit Touchscreen Unit to a USB port then select the fol lowing icon to download data from...

Page 16: ...n each case a drop down selection box will appear with the choice of setting 7 4 Analysis of B Scans PL Link provides a number of tools that assist in the analysis of B Scans Zoom into the B Scan usin...

Page 17: ...he various export options to be chosen In both cases the preview window shows the effects of the current out put selection Finish by clicking on export to select the file location name the file and in...

Page 18: ...Industrial Road 02 02A Morningstar Centre Singapore 536202 Phone 65 6382 3966 Fax 65 6382 3307 info asia proceq com Proceq Rus LLC Ul Optikov 4 korp 2 lit A Office 410 197374 St Petersburg Russia Phon...

Reviews: