ELECTROMAGNETIC COMPATIBILITY
Table 2
Guidance and manufacturer’s declaration - electromagnetic Immunity
The 240391 is intended for use in the electromagnetic environment specified below. The
customer or the user of the 240391 should assure that it is used in such an environment
Immunity Test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %
Electrical fast
transient/burst
IEC 61000-4-4
Power supply
lines
:
±2 kV
input/output
lines
:
±1 kV
Power supply
lines
:
±2 kV
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
line(s) to
line(s)
:
±1 kV.
line(s) to
earth
:
±2 kV.
100 kHz
repetition
frequency
line(s) to
line(s)
:
±1 kV.
100 kHz
repetition
frequency
Mains power quality should
be that of a typical
commercial or hospital
environment.