Introduction 

This manual should be used for initial set up of the system and saved for reference purpose. 

 

1.1

 

General Information 

This system has been tested and successfully approved to the following standards.   

EN60601-1 
EN60601-1-2 
 
 

 

EMC Warning Statement 

This equipment has been tested and found to comply with the limits for medical devices to the   
EN 60601-1-2:2004. These limits are designed to provide reasonable protection against harmful 
interference in a typical medical installation. This equipment generates, uses and can radiate radio 
frequency energy and, if not installed and used in accordance with the instructions, may cause harmful 
interference to other devices in the vicinity. However, there is no guarantee that interference will not 
occur in a particular installation. If this equipment does cause harmful interference to other devices, 
which can be determined by turning the equipment off and on, the user is encouraged to try to correct 
the interference by one or more of the following measures: 
- Reorient or relocate the receiving device. 
- Increase the separation between the equipment. 
- Connect the equipment into an outlet on a circuit different from that to which the other device(s) are 

connected. 

- Consult your homecare provider for help. 
 
This system has been tested and compliance to the following volunteer standards.   

„

 

FDA  

 

1.2

 

Intended use 

The Zzz-PAP Integrated Heated Humidifier is for use with the Zzz-PAP Mini-CPAP system. It has 
been designed to increase the humidity of the air from the CPAP therapy relieving the symptoms 
of dry throat and nasal passages that some people may experience.   
 

NOTE: Equipment not suitable for use in the presence of a flammable anesthetic 
mixture with air or with oxygen or nitrous oxide.

   

 
 
 
 
 
 

6

 

Service 

If the heated humidifier malfunctions or doesn’t work correctly for any reason, contact your 
homecare provider immediately. Never try to open the enclosure of heated humidifier. The heated 
humidifier requires servicing by qualified maintenance personnel.     

 
 

7

 

Troubleshooting 

The table below lists troubleshooting solutions for the problems that may happen. If the problem 
persists, contact your homecare equipment provider. 

 

Problem Possible 

Causes 

Solutions 

Power on indicator 
not lit 

1.

 

Power cord not plugged into a 
working AC outlet.   

2.

 

Power switch is not turned on   

3.

 

PCB or indicator worn out 

4.

 

Fuse or circuit breaker blown 

1.

 

Plug in the power cord to a working 
AC outlet.   

2.

 

See Setup procedure and turn the 
power on.   

3.

 

Contact your homecare provider. 

4.

 

Replace fuse, reset circuit breaker or 
contact an electrician. 

Heater on indicator 
will not light at all 

1.

 

heater plate over heated   

2.

 

PCB indicator worn out 

1.

 

Contact your homecare provider for 

assistance. 

2.

 

Contact your homecare provider for 
assistance. 

There is no air flow 
through the mask 

1.

 

CPAP device is not powered on or 
working correctly   

2.

 

Flexible tubing is not connected 
right.  

3.

 

Flexible tube is blocked.   

1.

 

Refer to your CPAP user manual.   

2.

 

Reconnect the flexible tubing 
correctly.  

3.

 

Unblock the flexible tube.   

Condensation in 
mask or flexible 
tube       

1.

 

The heater plate setting is too 
high.  

2.

 

The operating environment or 
position of heated humidifier is not 
acceptable - the room temperature 
is too low.   

1.

 

Adjust the control knob to lower 
temperature setting.   

2.

 

Remove any air conditioner which 
may nearby the heated humidifier. 
Or keep room temperature above 
57

° 

F. 

Water Leakage   

1

 

Assembly of water chamber is not 
correct.  

2

 

Water chamber or removable base 
plate is defective or broken.   

1.

 

Remove the water chamber from the 
heated humidifier, empty out the 
water and reassemble the water 
chamber again, make sure the 
removable base plate is aligned 
correctly to fit into water chamber, 
and fill as described in the manual. 
Test for leakage.     

2.

 

Replace the water chamber 
assembly.  

 
 

8

Summary of Contents for ZZZ-Pap

Page 1: ...rlboro New Jersey 07746 www pmiusa biz 666005 6130 V1 0 Integrated Heated Humidifier Owner s Manual Probasics Zzz PAP with Integrated Heated Humidifier Model No 9S 003 Series Item No 7501 Please read...

Page 2: ...RRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WARRANTIES IF ANY INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULR PURPOSE AND SHALL NOT EXTEND BEYONG THE DURATION...

Page 3: ...ANTY SHALL BE LIMITED TO SUCH REPAIR AND OR REPLACEMENT FOR WARRANTY SERVICE PLEASE CONTACT THE DEALER FROM WHOM YOU PURCHASED YOU PMI PRODUCT IN THE EVENT YOU DO NOT RECEIVE SATISFACTORY WARRANTY SER...

Page 4: ...ry therapist or physician z Never operate this product if a it has a damaged cord or plug b it is not working properly c it has been dropped or damaged d it has been dropped into water Return the prod...

Page 5: ...e power supply 2 Consult your homecare provider for other technical documents 9 NOTE CAUTION AND WARNING STATEMENTS DANGER Danger indicates an immediately hazardous situation which will result in deat...

Page 6: ...s or doesn t work correctly for any reason contact your homecare provider immediately Never try to open the enclosure of heated humidifier The heated humidifier requires servicing by qualified mainten...

Page 7: ...ems of the chamber are subject to normal wear and tear and may eventually be replaced Replace the chamber assembly as necessary 7 2 Product description Water Chamber The water chamber contains the wat...

Page 8: ...main outlet NOTE The plug is also serves as a disconnect device 5 4 Operation 4 1 Operating instruction Daily Use 1 Power on the heater by pressing the On Off switch on the rear of the heated humidif...

Reviews: