Thank you for purchasing a PRO-JECT AUDIO stylus balance. 

!

 

Warning of a hazard for the user, the unit or possible misuse.  

 

Important notice. 

Standard calibration 

1

 Turn on the meter with the 

ON/OFF

 button. Give your stylus balance sufficient warm up time. Usually 30 

seconds to give the internal components time to stabilise before calibrating. 

2

 Press and hold the 

TARE

 button until the display shows 

CAL

3

 Press the 

TARE

 key again. Now the display shows (flashing) 

CAL

 and 

5.000

4

 Gently place the 5g weight on the centre of the platform. After approximately 5 seconds the display will 

read (flashing) 

PASS

 and 

5.000

 

Occasionally the display will show for example 

5.002

 or 

5.004

 instead of 

5.000

. This is no defect. 

It will not affect the calibration.

 

5

 Remove the weight an press 

ON/OFF

 to turn off the meter.

 

Cartridge downforce adjustment

 

1

 Remove the stylus guard and the stylus balance protection cover. Gently place the meter on the platter. 

Move the headshell over the weighting platform. 

2

 Turn on the stylus balance with the 

ON/OFF

 button, the display shows 

0.000

. Give your meter sufficient 

warm up time. Usually 30 seconds to give the internal components time to stabilise. 

3

 Gently lower the tonearm until the stylus is on the black dot of the weighting platform. The display shows 

the tracking force. 

4

 Lift the tonearm and adjust the counter weight (tracking weight dial) and repeat the above procedure the 

recommended tracking force is achieved. 

!

 

Never load the stylus balance with more than 

5g

, overloading can permanently damage it.

 

Special calibration - 

if the display shows 

OUTZ

 after turning on

 

1

 Turn off the meter with the 

ON/OFF

 button. 

2

 Press and hold the 

TARE

 button, the immediately press the 

ON/OFF

 key once until 

CAL

 will be displayed. 

After releasing the 

TARE

 key, the display shows a random number. 

3

 After pressing the 

TARE

 button again, the display will read 

ZERO

 and another random number. 

4

 Gently place the 5g weight on the centre of the platform, wait for 2 seconds and press the 

TARE

 key again. 

The display will read (flashing) 

PASS

 and 

5.000

 

Occasionally the display will show for example 

5.002

 or 

5.004

 instead of 

5.000

. This is no defect. 

It will not affect the calibration.

 

5

 Remove the weight an press 

ON/OFF

 to turn off the meter.

 

Other display readouts 

No display, display shows 

LO

 or 

8888

 - batteries are empty and need to be replaced. 

O_LD

 - overload (more than 5g). Please remove the excessive load immediately. 

LLLLL

 - The weighting load cell has been damaged by misuse (i.e. it was shocked, dropped or overloaded). 

You can try to recalibrate (standard calibration) the meter. If the load cell has not been damaged to much, it 
will work again after recalibration. 

UNST

 - Use the meter on a more stable, vibration free surface. 

Backlight 

The 

LIGHT

 key switches the backlight off an on. 