Prizmatix
10 |
P a g e
UHP-F LED Illuminator User Manual
4.
Connect the other end of the Light Guide to the microscope port by appropriate
adaptor (sold separately).
5.
Switch the back-panel power switch to the "ON" position. The LED light should be
seen at the microscope slide.
CAUTION!
:
Do not stare at operating lamp. May be harmful to the eyes
6.
In case you are using digital camera to inspect the sample adjust the camera
exposure to get best image.
7.
At end of use switch the unit off by switching the Power Switch on back panel to
OFF position.
CAUTION
!:
Never disconnect the power cord form the product before
switching the ON/OFF switch on back panel to OFF position
8.
For TTL input connect BNC cable to
TTL
input connector on back of unit. To enable
TTL control change the position of the
TTL
Int/Ext
switch to
Ext
position.
9.
For Analog Input connect BNC cable to
Ain
input connector on back of unit. To
enable Analog Input control change the position of the
Ain
Int/Ext
switch to
Ext
position
CAUTION!
:
Do not cover back panel of unit - ensure that air can circulate
freely.