Professional ultrasonic nebulizer
Troubleshooting
Trouble
Where to inspect
How to correct
Insufficient amount of
water
Is the water in the water tank up
to the level float?
Fill the water up to the level
float.
The unit does not
nebulize.
Are the air vents at the
bottom blocked?
Unblock the air vents at the
bottom of the unit. Wait for the
unit to cool down.
Is the Isapak® chamber
correctly set?
Set the Isapak® chamber
correctly.
Is the amount of medication in
the Isapak® chamber adequate?
Change the medication bottle.
Is the inhalation hose bent and
the medication pooled in the
hose?
Adjust the bent hose
correctly and drain the
pooled medication.
Is the aeration hose
placed correctly?
Place the aeration hose correctly.
The amount of aerosol has
decreased.
Is the room temperature too
low and/or water temperature
too low?
Operate the unit without
medication for about 4
minutes, then use the unit.
Is the bacterial filter blocked?
Replace the bacterial filter
with a new one.
Nebulization is unstable.
Is the amount of
medication exceeding 150
ml?
Reduce the amount of
medication to less than 150
ml.
• If the unit does not operate normally after taking the above-mentioned measures, do
not touch the internal mechanism and consult the store where you purchased the unit
or PRIZMA service centre.