Cleaning

 

Please follow the cleaning guidelines below on cleaning and disinfecting the Gantry. 

 

5.1

 

General cleaning 

It is recommended to clean the FSG200 and accessories before the use by a different patient, reducing the risk of 
cross

contamination. 

 

The exterior of the FSG200 can be cleaned using a damp soapy cloth for general cleaning duties. Please ensure the cloth is damp 
and not wet. Ensure the exterior of the device is dry after cleaning. Dry using a clean dry cloth. 

5.2

 

Disinfecting (if necessary) 

Should the Gantry require a more thorough clean, the use of the Actichlor

 disinfectant product (which is widely available in 

tablet form and used throughout the health care industry) is recommended. 
 

Follow the manufacturer’s safety instructions for the use of the cleaning 

product before use to ensure safe use for 

the operator and the patient. 
Ensure the cloth is damp before the cleaning process.  

 
 
 
Application is through a clean damp cloth applied to wipe the device down. Use in the following dilutions to ensure an effective 
clean: 

 

Actichlor

 dissolvable chlorine tablets provide a concentration of 1000 ppm of available chlorine (0.1%) per 1 tablet 

 

1 tablet (1.7g formed tablet (x1)) will create a virucidal solution, diluted in 1 litre of water to provide effective means to 
c

lean a “dirty” device. This is also ideal for use after an outbreak of the Norovirus/winter vomiting and can be used as a 

precaution against C.Diff. It is effective against viruses, bacteria, spores, yeasts and moulds. 

 

The contact time against the outer components of the device should be for 5 minutes to prevent any virucidal 
infections without a degradation to the functionality of the device. 5 minutes is a recommended contact time. The 
device can withstand a longer contact period but the 5 minute recommendation as a minimum must be followed to 
provide an effective cleaning regime. 

 

Blood spills should be dealt with by an increased concentration of the solution 

 please refer to the instructions on the 

manufacturers product labelling. 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 