Connect the NPC-D-6110 controller to the computer using an RS232C cable from the serial port on 
the computer to the SYNC IN/OUT port on the NPC-D-6110.

  

A gender changer adaptor is required 

to connect the RS232C cable to the sync port. The adaptor required is a 9-pin male to male adapter 
as shown.   

On ‘power up’, the controller will always sweep the stage across its range to

 

carry out auto-

calibration. 

It is important to ensure that there is sufficient clearance between the lens, sample 

and illumination to allow this to take place.  

If the temperature or load changes significantly  

the stage can 'clip' at one end of the travel.  Should this happen restart your system to allow  
auto-calibration. 

 

Once installed navigate to Device Manager and make a note of the serial COM port used. 
Open NIS-Elements and navigate to the manage device section; install devices as shown below and select “Prior ProScan III/ 
OptiScan III”.

 

 

 Set the port to the setting found under Device Manager. 

CONNECTING THE CONTROLLER TO THE COMPUTER

 

NIS-ELEMENTS

 

9 pin male to male 

adaptor

 