background image

8

  

 

Prior 

Scientific 

9. Phase 

Contrast 

 

 

  

   

 

 

 

 

 

 

      

Figure 

Phase contrast is a technique used to amplify contrast in unstained samples.  To be able to 

see objects with phase contrast the optional phase kit is required.  This includes a phase 

turret condenser (see figure 2), phase objectives, a centering telescope and a green filter.  

Firstly the standard condenser must be removed by loosening the retaining screw and the 

phase turret condenser installed in its place.  Swap the standard objectives (if you have 

them) for phase contrast objectives, these are engraved PH, and select the correct condenser 

position (0 for brightfield, 10 for the 10x objective, 20 for the 20x objective etc.)  Follow 

the steps in section 7 to ensure that Köhler illumination is set up properly. Remove an 

eyepiece and insert the centering telescope, this allows you to focus on the phase rings and 

make adjustments to their alignment.  The two rings, one bright, one dark should overlap 

perfectly (A), if not (B) then their position relative to one another should be adjusted using 

the centration screws at the rear of the condenser (Figure 3). 

To make adjustments to the phase ring alignment push in the adjustment screws and then 

turn. When the alignment is complete remember to pull the screws back out or it will be 

impossible to change condenser position. 

Summary of Contents for PriorLab

Page 1: ...PriorLux 1...

Page 2: ...king The PriorLab PriorLux is shipped in protective bags within a pre formed container Each component should be carefully unpacked and checked cutting rather than tearing the plastic bags The head bin...

Page 3: ...sion control and focus stop Co axial fine and coarse adjustment with tension control and focus stop Viewing Head Interpupillary distance 55 75mm Interpupillary distance 55 75mm Kohler Illumination 12V...

Page 4: ...4 Prior Scientific 4 Component Parts...

Page 5: ...ng on the power reduce the lamp intensity control to its minimum setting After switching on the intensity should be increased to a comfortable level This procedure should be reversed before switching...

Page 6: ...ct centration Figure 1 Pictures C D E Fully open the field diaphragm F Fully open the condenser diaphragm and then slowly close it until you see the contrast within the image increase If you then remo...

Page 7: ...the objective does not touch the slide Remove residual oil from the objective after use 8 Cleaning Objectives It is critical that the front lens of each objective is kept clean and free of contaminat...

Page 8: ...d select the correct condenser position 0 for brightfield 10 for the 10x objective 20 for the 20x objective etc Follow the steps in section 7 to ensure that K hler illumination is set up properly Remo...

Page 9: ...trast image is poor then check that the condenser is properly focused and centred and then ensure that the rings are correctly aligned 10 Dark Ground The microscope may be used for dark ground observa...

Page 10: ...the c mount adapter and insert the adapter with the attached camera into the top of the photo tube on the trinocular head C Tighten the screw to secure the assembly D Connect camera to a PC framegrab...

Page 11: ...PriorLux 11...

Page 12: ...its back to gain access to the base plate D Loosen the screw which holds the lamp cover to the base plate E Open the lamp cover to expose the bulb holder F Remove the old bulb and replace it with the...

Page 13: ...following symbols have been used on this microscope These symbols are found next to the bulb access door on the underside of the instrument Warning High voltage disconnect power supply before changing...

Page 14: ...use part 1 General requirements EN61326 1997 A1 A2 A3 Electrical equipment for measurement control and laboratory use EMC requirements Class B emmisions EN61326 1997 A1 A2 A3 Electrical equipment for...

Page 15: ...PriorLux 15 17 Specification subject to change without notification Part No W2968 Issue B 10 05...

Reviews: