Princeton PHV-400 Operating Instructions Download Page 3

PHV-400.DOC 

control where fully clockwise is –1250 

V

olts DC.  The output voltage can be 

monitored on the digital meter as it is adjusted.  The programming voltage has no 
effect in the LOCAL mode.  It is advisable to adjust the front panel control fully 
counter clockwise before changing the position of the REMOTE/LOCAL switch. 
 

 

Specifications: 

Output 

V

oltage 

 

 

 

0 to –1250 

V

olts DC 

Output 

C

urrent 

 

 

 

2 m

A

 maximum 

Programming 

V

oltage 

 

 

0 to +5 

V

olts DC for o to –1250 

V

olts DC output 

Accuracy 

 

 

 

 

  

+/- (3% of s 0.5% of full scale) 

Line Regulation 

 

 

 

  

+/- 0.001% for input variation of +/-1% 

Ripple 

 

 

 

 

  

0.001% peak-peak at maximum power output 

Stability (after 30 minute warm-up) 

  

0.005% per hour, 0.02% per 8 hours 

Input 

C

onnector 

 

 

 

        

Switchcraft #142A mating plug #750 

Reviews: