13.3.3 Gating Mode
LightField supports three Gating Modes for RF experiments:
•
Repetitive
;
•
Sequential
;
•
Custom
.
The desired option is selected from the
Mode
drop down list on the
SuperSYNCHRO
Timing
Figure 13-2: SyperSYNCHRO Timing: RF Mode Options
Each mode has its own set of configuration parameters which are described in the following
sections.
13.3.3.1 Repetitive
Perform the following procedure to configure Repetitive Gating Mode:
1.
The first requirement is that the PI-MAX4: 1024i-RF camera be aligned and focused on
the area of interest in the experiment. This is best accomplished while the camera is
operating in Full Frame readout mode. Verify that the
Phosphor Decay Delay
is
appropriate to the phosphor used by your camera: the phosphor decay delay time
entered in LightField can be viewed or changed after clicking on the
Advanced
button
on the
Common Acquisition Settings
expander. The procedure for initial focus is
.
NOTE:
The
Phosphor Decay Delay
setting tells LightField how long to
wait after the gate pulse to shift the image. If there is some
residual image from the first frame in the second frame, simply
increase the
Phosphor Decay Delay
setting to allow more time
for the phosphor emission to decay before shifting the image.
2.
After the alignment and focus, the PI-MAX: 1024i-RF system needs to be put into RF
modulation mode. Configure the following parameters on the
Common Acquisitions
Settings
expander
•
Select
Use RF Modulation
.
4411
-013
9_018
8
Summary of Contents for PI-MAX4
Page 1: ...PI MAX4 Camera System 4411 0139 Issue 9 July 3 2018 ...
Page 4: ...4 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 18: ...18 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 32: ...32 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 46: ...46 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 96: ...96 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 156: ...156 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 172: ...172 PI MAX 4 System Manual Issue 9 ...
Page 216: ...216 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 248: ...248 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 258: ...258 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 274: ...274 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 292: ...292 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 296: ...296 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 304: ...304 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 310: ...310 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 326: ...This page is intentionally blank ...