6. Pull the retaining plate up slightly and then forwards until it loosens and can be removed easily.
7. Carefully remove the measuring tip.
18.5.3
Assemble the measuring tip
1. Insert the new measuring tip (note that the inlet aperture is located on the convex side, see Fig. 18.3 on
page 53) or turn the measuring tip around.
Pinhole
Fig. 18.3:
Inlet aperture (pinhole) in the measuring tip
2. Use up and down movements to insert the retaining plate into the rotational disc at an angle of about
45 degrees, ensuring that the guide groove is pointing upwards. Then push it down into the recess
(see Fig. 18.4 on page 53).
3. Check whether the retaining plate is positioned correctly by lightly pressing the front edge.
Fig. 18.4:
Inserting the retaining plate
4. Insert the fastening screws (Torx T8) and tighten manually.
If you have turned the measuring tip, use the arrow buttons to select the “Measuring Tip” setting in
LaserDiagnosticsSoftware LDS in the menu
Device control > Advanced > Measuring tip
.