![Primes BeamMonitor BM+ Series Operating Manual Download Page 37](http://html1.mh-extra.com/html/primes/beammonitor-bm-series/beammonitor-bm-series_operating-manual_1601730037.webp)
BeamMonitor BM+
35,0(6
37
Revision 02/2019 EN
17.3.3
Assemble the detector
Only use isolating plastic screws to fasten the detector to prevent any electric noise signals from
being interspersed. Do not forget the foam rubber spacer during installation, otherwise the rota-
tional disc could be mechanically blocked by the screws. The foam rubber spacer also ensures
mechanical decoupling of the detector.
1. Place the foam rubber spacer (C) on the mounting surface of the detector (see Fig. 17.5 on page 37).
2. Connect the cables.
NOTICE
Blocking the rotational disc
If the screws are tightened too firmly, they might block the rotational disc!
X
Only tighten the screws hand-tight. The foam rubber spacer must only be compressed to no
more than half its original thickness.
3. Fasten the detector with the two plastic screws (D).
C
D
Fig. 17.5:
Assembling a new detector
17.3.4
Replace cover
1. Place the cover on the casing (see Fig. 17.1 on page 35).
2. Screw the cover on tight with the four torx screws T8.
3. Check that the cover is securely seated.
• The cover must lay flush with the casing.