TROUBLESHOOTING GUIDE
SYMPTOM
POSSIBLE ISSUES OR COMMENTS
SOLUTIONS
CPR Rate Monitor
Cannot find CPR Rate
Monitor (Monitor) – no
lights in the diaper area
You have purchased/received a Prestan
Professional Infant Manikin without the
CPR Rate Monitor
If you were expecting a monitored
manikin, check with the distributor
where you purchased the manikin to
insure you received the item you paid
for. See the chart on page 2 of this
manual for the product numbers.
Note: Infant CPR Rate Monitors may be
purchased separately from your
Prestan Authorized Distributor. They
are shipped with a replacement diaper
area cover for viewing the lights
through it.
CPR Rate Monitor
(Monitor) lights are not
working properly
Please follow the
troubleshooting tips at
right in the order they are
listed for best results
Chest compressions are not deep
enough to signal Monitor
Batteries are not installed, need
replacing or are not inserted correctly
in the CPR Rate Monitor
Either the Clicker or the Monitor needs
to be replaced
Compress the chest until the click is
audible from the Clicker at 2”-2.4”
Install or replace batteries into Monitor
(front of manikin, top of diaper) –
requires 2-AA (1.5V) batteries
Exchange Monitors with working
manikins (if available) to confirm
whether the Monitor
(RPP-IMON-1) needs to be replaced*
CPR Rate Monitor
(Monitor) lights are all on
Batteries are running low
Replace batteries
CPR Rate Monitor
(Monitor) red light is
blinking
This is normal after replacing batteries
or manikin is paused - may blink up to
16 times to indicate ready status
Commence or resume compressions -
blinking will stop and feedback rate
signals will begin
CPR Rate Monitor falls out
when administering back
blows
Monitor cover in diaper area is no
longer fitting properly
Request new Infant Monitor Cover
(10402)*
*
Contact your Prestan Authorized Distributor to either purchase the replacement part or receive it at no charge if the
manikin is within Prestan’s 3-year warranty period and is a qualifying warranty repair.