FEATURE SETTINGS
Ventilations
On
Ventilations
Off
CP
R
M
et
ro
no
m
e
O
n
1. Program begins with analyzing
patient and advising shock or no
shock based on user scenario
setting.
2. Audible prompt “BEGIN CPR” starts
compression cycle.
3. Metronome beats and audible
prompt notes “TEN”, “TWENTY”, and
“THIRTY” compression counts.
4. After thirty (30) compression
tones, the metronome pauses
for an audible prompt to “GIVE 2
BREATHS”.
5. Compression tones and breaths
continue for five (5) cycles (two (2)
minutes).
6. Unit analyzes patient.
7.
Cycles repeat indefinitely following
scenario shock pattern until unit is
manually shut off.
1. Program begins with analyzing
patient and advising shock or no
shock based on user scenario
setting.
2. Audible prompt “BEGIN CPR” starts
compression cycle.
3. Compression tones with audible
prompt to “CONTINUE CPR” every
twenty-four (24) seconds (x4).
4. Compression tones continue for two
(2) minutes.
5. Unit analyzes patient.
6.
Cycles repeat indefinitely following
scenario shock pattern until unit is
manually shut off.
CP
R
M
et
ro
no
m
e
O
ff
1. Program begins with analyzing
patient and advising shock or no
shock based on user scenario
setting.
2. Audible prompt “BEGIN CPR” starts
compression cycle.
3. Audible prompt notes “TEN”,
“TWENTY”, and “THIRTY”
compression counts.
4. After time for thirty (30)
compressions, the metronome
pauses for an audible prompt to
“GIVE 2 BREATHS”.
5. Compression tones and breaths
continue for five (5) cycles (two (2)
minutes).
6. Unit analyzes patient.
7.
Cycles repeat indefinitely following
scenario shock pattern until unit is
manually shut off.
1. Program begins with analyzing
patient and advising shock or no
shock based on user scenario
setting.
2. Audible prompt “BEGIN CPR” starts
compression cycle.
3. Silence except for audible prompt to
“CONTINUE CPR” every twenty-four
(24) seconds (x4).
4. Cycle stops at two (2) minutes.
5. Unit analyzes patient.
6.
Cycles repeat indefinitely following
scenario shock pattern until unit is
manually shut off.