Page 2: ...E GENERAL INFORMATION Use this instruction sheet as a reference for the Prestan Professional AED Trainer PLUS and Accessories Please read through instructions before first use Warning The Prestan Professional AED Trainer PLUS is intended for training purposes only It is NOT an Automated External Defibrillator AED and will not deliver a shock To avoid confusion it should NEVER be kept in the same a...
Page 3: ...ilable sold separately Controls multiple units Long Battery Life Low battery voice prompt warning Automatic shut off for extended battery life Programming is compliant with current AHA Guidelines 2015 Easy to replace language module if guidelines change HIGH PERFORMANCE TRAINING PADS Start by installing your training pads onto the unit s pre installed cables See page 6 for full instructions Patent...
Page 4: ...S SHOCK BUTTON CHILD BUTTON FOUR 4 C BATTERIES PRE INSTALLED IN UNIT TRAINING PADS STORAGE CASE ON OFF PAUSE BUTTON PRODUCT INFORMATION LABEL MODEL SERIAL NUMBER BATTERY COMPARTMENT NON MARKING SILICONE PROTECTIVE FOOT 4 SPEAKER CONTROL PANEL LANGUAGE MODULE FRONT OF UNIT BACK OF UNIT TRAINING PADS CONNECTOR ON OFF STATUS INDICATOR LIGHTS ...
Page 5: ...tor light above the Child Button will light up when the Child option has been selected and is active Indicator light turns off if the program is paused but turns back on when the program is resumed SHOCK BUTTON Red Shock Button will flash when program instructs to deliver shock In Semi Automatic mode pushing the Shock Button while it is flashing will simulate delivering a shock and will stop the f...
Page 6: ... corner 6 Pad Assembly a Open the plastic package and remove the Prestan Professional Training Pads b Find the ends of the AED Trainer cables that are not connected to the unit c Insert each pad into a cable connector pad tab may need to be wiggled into place if fit is snug and line up the hole in the tab of the pad with the hole in the connector Ensure pads are not installed upside down see step ...
Page 7: ...tors 5 Refer to Step 6 Pad Assembly on previous page to complete the installation f Open the white plastic pad carrying case found in the unit g Turn pad over and remove the clear plastic liner from the AED Pad by gently pushing down on the tab end of the pad causing the liner to separate h THROW THE CLEAR PLASTIC LINER AWAY re using the liner does not extend the life of the adhesive and may impac...
Page 8: ...INDICATOR LIGHTS The SELECTION INDICATOR LIGHTS are below the Language Selection button The SELECTION INDICATOR LIGHTS are used during the setup of the unit They indicate settings such as automatic or semi automatic shock mode ventilations modes and metronome Indicator lights are read left to right LANGUAGE SELECTION BUTTON PAUSE PLAY BUTTON VOLUME BUTTON SCENARIO SELECTION BUTTON SELECTION INDICA...
Page 9: ... the five 5 available scenarios Scenario 1 shock no shock no shock Scenario 2 shock no shock shock Scenario 3 shock shock no shock Scenario 4 no shock no shock no shock Scenario 5 shock shock shock VOLUME BUTTON There are three 3 volume levels on the unit By pressing the VOLUME BUTTON you can increase or decrease the volume of the unit depending on need The indicator lights will light up to indica...
Page 10: ...and save press the On Off Button PROGRAMMING OPTIONS VENTILATIONS ON OFF The Prestan Professional AED Trainer PLUS can allow time for administering breaths between compression rounds This option can be toggled on or off The AED Trainer is preset from the factory with ventilations ON To change this mode With the AED Trainer off hold down the On Off Button until two 2 beeps are heard about 7 10 seco...
Page 11: ...r off hold down the On Off Button until two 2 beeps are heard about 7 10 seconds then release the first beep is for test mode so wait for the second beep the Shock Button will be flashing on and off Look at the Selection Indicator Lights there will be three 3 lit up in red or yellow depending on your Automatic Semi Automatic preference Press the Scenario Selection Button below the lights twice unt...
Page 12: ...y four 24 seconds x4 4 Compression tones continue for two 2 minutes 5 Unit analyzes patient 6 Cycles repeat indefinitely following scenario shock pattern until unit is manually shut off CPR Metronome Of f 1 Program begins with analyzing patient and advising shock or no shock based on user scenario setting 2 Audible prompt BEGIN CPR starts compression cycle 3 Audible prompt notes TEN TWENTY and THI...
Page 13: ... If it becomes soiled simply wipe with a clean slightly damp soft cloth If the Trainer is extremely dirty common cleaners can be used cautiously DO NOT USE ABRASIVE MATERIALS OR CLEANING PRODUCTS ON THE UNIT DO NOT SUBMERGE THE UNIT IN WATER TROUBLESHOOTING TEST MODE Is the unit programming not progressing past place pads on patient Enter TEST MODE to determine if your pads or cables need to be re...
Page 14: ...d around and re secure with black connector pin see Training Pad Replacement instructions on Page 7 Replace pad pads last approximately 100 uses with proper care Device proceeds to analyzing before both pads have been placed on the manikin Pads not properly positioned in pad case prior to starting program Re position pads in case to lay completely flat and sensor is positioned in recessed area of ...
Page 15: ...DT2 CABLE Set of Replacement Cables for the Professional AED Trainer PLUS PP ACASE2 1 Replacement Pad Storage Case for the Professional AED Trainer PLUS PP AEDT2 101 M English Spanish PP AEDT2 102 M English French Language Module for the Professional AED Trainer PLUS Additional languages available based on distributor represented geographic area PP AEDT2 100 R Remote Control for the Professional A...