Measuring arrangements

4.1

The front of the instrument (Fig. 2)

Fig.2.

Establish a ground connection between ground of the laboratory and the ground terminal

of the instrument.

Because the instrument is built according to safety class I of IEC 359, the mains cable
contains a ground conductor which also has to be connected to ground of the local mains.

It is strictly forbidden to open the ground connection between "labo-

ratory ground

" and "Instrument

ground of local mains"!

Danger of life!

4.2

Using the instrument with a divider.

HV to be measured

CH

CL

1M

KV-M100

BNC

Front

Back

H

L

S

u

rg

e

A

rr

e

s

te

r

2

3

0

V

Surge Arrester 90V

The instrument can be used with resistive, capacitive, inductive divider or voltage trans-
former.