Establish a ground connection between ground of the laboratory and the ground terminal
of the instrument.
Because the instrument is built according to safety class I of IEC 359, the mains cable
contains a ground conductor which also has to be connected to ground of the local mains.
It is strictly forbidden to open the ground connection between "labo-
ratory ground
" and "Instrument
ground of local mains"!
Danger of life!
4.2
Using the instrument with a divider.
HV to be measured
CH
CL
1M
KV-M100
BNC
Front
Back
H
L
S
u
rg
e
A
rr
e
s
te
r
2
3
0
V
Surge Arrester 90V
The instrument can be used with resistive, capacitive, inductive divider or voltage trans-
former.