PREPLINE PSP20 User Manual Download Page 4

 

 
 
6.4 For setting a new reference weight, press and hold the „PCS” -key until the display starts 

blinking, then continue as from step 1. 

    
    
    

Calibration

Calibration

Calibration

Calibration    

    

To perform the calibration, you need exactly standard weight. 
 
1.1 When the scale is switched off press and hold the “MODE“ key. 
1.2 Additionally press shortly the “ON/TARE“ key, the display shows a number. 
1.3 Release the “MODE“ key. 
1.4 Press again the “MODE“ key, the display shows „5000“ 
1.5 Place a 5 kg calibration weight on the scale, the display now shows “10000“ 
1.6 Place a 10 kg calibration weight on the scale, next “PASS“ appears on the display and 

finally the scale shows the normal weighing display.   

1.7 The scales is now recalibrated.   

If in any case the procedure should have failed, the calibration should be repeated.   

 
Note: During recalibration, the scales should not experience any movement or draught! 

 
 
 

Explanation of  the special symbols

Explanation of  the special symbols

Explanation of  the special symbols

Explanation of  the special symbols    
    

1.1. Switch-On 

After pressing the “ON/TARE“–key all symbols appear. One can check if all segments are 
shown  correctly.  The  “zero”  which  then  appears  shows  that  the  scales  is  ready  for 
weighing. 
 

1.2 Negative Weight Display 

Press the “ON/TARE“–key again. 
 

1.3 Overload 

If  the  weight  on  the  scales  is  heavier  than  the  max.  capacity  of  the  scales,  then 
“O-ld“ appears in the display. 
 

1.4 Power Supply 

“Lo“ means that the battery is empty and needs to be replaced. 