18
100~165
Normal
3. Apply control solution
Shake the control solution vial thoroughly before use.
Squeeze out a drop and wipe it off, then squeeze out another drop and
place it on the tip of the vial cap.
Hold the meter to move the absorbent hole of the test strip to touch the
drop. Once the confirmation window fills completely, the meter will begin
counting down.
To avoid contaminating the control solution, do not directly apply control
solution onto a strip.
4. Read and Compare the Result
After counting down to 0, the control
solution test result will appear on the
display. Compare this result with the
range printed on the test strip vial and it
should fall within this range. If not, please
read the instructions again and repeat
the control solution test.
" The blood glucose is (number)
milligrams per deciliter. "
Premium V12 Voice model only.
Voice Guide: