background image

 

For  example,  the  measured  average  temperature  of  the  heating  surface  is  30

,  but  the  LED 

temperature display shows 25

.  Use the “▲” button to add 5

 (the difference in temperature 

values). Then when you exit the settings selections (step 5) the LED temperature display will 

read 30

 

2)

 

Press the “SET” button again and “E” will be shown in the LED temperature display. Press 

the “SET” button again and “00.5” will be shown. 

DO NOT CHANGE THIS VALUE

. It is 

a factory preset and changing it will cause the Slide Warmer to malfunction.

 

 

3)

 

Press the “SET” button again and “P” will be shown in the LED temperature display. Press 

the “SET” button again and “15.0” will be shown. See step 6 for instructions on when to 

adjust this value.

 

 

4)

 

Press the “SET” button again and “c” will be shown in the LED temperature display. Press 

the “SET” button again and “10.0” will be shown. 

DO NOT CHANGE THIS VALUE

. It is 

a factory preset and changing it will cause the Slide Warmer to malfunction.

 

 

5)

 

Press the “SET” button again and your settings will be saved.

 

 

6)

 

If the Slide Warmer does not maintain its recalibrated temperature: 

Press and hold the “SET” button for 3 seconds until “Sc” is shown. Then press the “SET” button 

4 more times and “P” will be shown in the LED temperature display. Press the “SET” button 

again to see the value of “P”; this is the temperature control value. This value is equal to P/10°. 

This value can be adjusted so that the heating surface maintains a steady temperature. Use the 

“▲” or “▼” buttons to set the value between “0.00” and “99.9”. 

 

For example, if the LED temperature display value is “50.0” and the value of “P” is “10.0” the 

heater will turn off and on at P/10 = 10/10 = 1°C.  So the heater will turn off at 51°C and heat 

back up if it falls below 49°C. 

If the temperature rises too slowly or it cannot reach the “SET” temperature increase the value of 

“P”. If the temperature rises too fast or it surpasses the “SET” temperature decrease the value of 

“P”. 

 

7)

 

Press the “SET” button again and the LED temperature display shows “c” (see step 4). 

 

8)

 

Press the “SET” button again to save your settings. 

 

9)

 

Monitor for 10 minutes and readjust values as necessary.  

 

 

 

 
 

Summary of Contents for XH-2001

Page 1: ...in the best results of this equipment Warning To prevent fire or shock hazard do not expose this unit to rain or moisture Unless labeled otherwise this equipment should be used with AC 120V 60Hz in th...

Page 2: ...1 1 1 Table of Contents Section Table of Contents Page 1 Safety Page 2 Contents Specifications Page 3 Control Panel Setup Page 4 Operation Page 5 Troubleshooting Page 5 6 Warranty Page 7...

Page 3: ...ing the fuse 7 If a problem occurs with the device contact your dealer Individuals that are not qualified to repair this device should not attempt to do so the result could be additional damage that w...

Page 4: ...H 2001 Large XH 2002 Small Overall size 25 x 9 x 4 640 x 230 x 100mm 10 x 8 x 3 260 x 200 x 85mm Surface size 25 x 8 10 x 7 Capacity Approx 66 slides Approx 23 slides Power 200 Watts 100 Watts Net Wei...

Page 5: ...display 4 Heat indicator lamp 4 2 Setup 1 Open box and remove slide warmer from packaging 2 Place slide warmer onto flat surface 3 Plug AC power cord firmly into back of unit connect plug to power sup...

Page 6: ...1 Troubleshooting Calibration testing Turn the Slide Warmer on and allow it to warm up for 15 minutes Then use a surface or infrared thermometer to measure the average temperature between the edge of...

Page 7: ...Press the SET button again and your settings will be saved 6 If the Slide Warmer does not maintain its recalibrated temperature Press and hold the SET button for 3 seconds until Sc is shown Then pres...

Reviews: