7.1 Operations

 

 
Time Adjustment 

1.

 

Press “Switch” to alternate to time setting. Push  the “+” or “-” button to adjust the 
time setting. To increase the run time push the “+” button until desired time reached 
or maximum of 60 minutes. (See section 6.1). To decrease the run time push the “-” 
button until desired time reached or minimum of 1 min. (See section 6.1). If the 
numbers above “TIME LED” are not Lit, make sure  AC  power cord is completely 
plugged into socket.  

 
 
Speed Adjustment 

1.

 

Push the “SWITCH” button one time to adjust the speed setting.  The 4 digit number 
beside the word “TIME/SPEED LED” should change to rpm’s.  (See section 6.1). If 
the numbers above “TIME/SPEED LED”  do not change to rpm’s, then continue to 
push the “SWITCH” button until this happens.  

2.

 

This centrifuge default setting for speed is in RPM’s.  Please select desired speed 
(shown in RPM’s) by using the “+” or “-” arrow buttons until the display panel reads 
your desired settings. 

 

 

 

 

Spinning Centrifuge 

2.

 

Insert the centrifuge tubes in a manner that will balance the specimen tubes’ weight.  
Always insert an even number of tubes on opposite sides of the rotor with 
approximately the same amount of specimen in each tube. 

 
Now that the desired time is selected, you can now run your samples. 

1.

 

Make sure lid is closed and locked.  

2.

 

Press the “On/Off” button, and centrifuge will begin spinning. 

3.

 

After the centrifuge has run for the programmed  amount of time unit will begin to 
decelerate and come to a complete stop.  Please note that the  lid should never be 
opened until centrifuge has come to a complete stop. 

 

 

 
Note 

This centrifuge lid is designed to automatically stop when the lid is open.  If you have any 
problems operating the centrifuge; Please contact your dealer or authorized provider. 

 

Summary of Contents for XC-2000

Page 1: ...in the best results of this equipment Warning To prevent fire or shock hazard do not expose this unit to rain or moisture Unless labeled otherwise this equipment should be used with AC 110V 60Hz in th...

Page 2: ...1 Table of Contents Section Table of Contents Page 1 Safety Page 2 Contents Accessories Page 3 Specifications Page 4 Notes of Operations Page 5 Control Panel Setup Page 6 Operations Page 7 Warranty Pa...

Page 3: ...under the maximum speed when the density of the sample exceeds 1 2g cm the maximum speed of the rotor must be reduced accordingly When the centrifuge is running rotor is spinning Please keep an area o...

Page 4: ...ne tube cushions 6pc 4 Fuse 1pc 4 AMP 5 Operating manual 1pc Upon receipt please examine the package if there s anything wrong please kindly contact your dealer promptly 3 2 Accessories If you need op...

Page 5: ...ly Setting Time Range 1 60min Maximum RPM 4000 r min Maximum relative centrifugal force RCF 1000 rpm 112 RCF 1500 rpm 252 RCF 2000 rpm 447 RCF 2500 rpm 699 RCF 3000 rpm 1006 RCF 3500 rpm 1370 RCF 4000...

Page 6: ...en lid and remove specimen tubes Use only with accessories provided or parts designed specifically for use with this model Do not use centrifuge when the door is open Make sure the specimen tubes are...

Page 7: ...nate display between time speed On Off run stop 6 2 Setup 1 Open box and remove centrifuge from packaging 2 Place centrifuge onto flat surface 3 Open lid and remove foam packaging inside 4 Plug Power...

Page 8: ...etting for speed is in RPM s Please select desired speed shown in RPM s by using the or arrow buttons until the display panel reads your desired settings Spinning Centrifuge 2 Insert the centrifuge tu...

Page 9: ...air technicians or damage occurring in transit If you have questions concerning this product or warranty contact the dealer from whom it was purchased For warranty service product should be well packe...

Reviews: