premiere XC-2000 Operating Manual Download Page 1

 

Centrifuge 

 

Model XC-2000 

Operating Manual 

 

Before using this product, please read this manual carefully to learn about the available 

features to obtain the best results of this equipment. 

 

Warning

:

 To prevent fire or shock hazard, do not expose this unit to rain or moisture. Unless 

labeled otherwise, this equipment should be used with AC 110V 60Hz in the USA or Canada. 

220V 50Hz available as Model No.

 XC-2000-220 

Caution

No user-serviceable parts inside. Refer servicing to qualified service personnel

 

 

Summary of Contents for XC-2000

Page 1: ...in the best results of this equipment Warning To prevent fire or shock hazard do not expose this unit to rain or moisture Unless labeled otherwise this equipment should be used with AC 110V 60Hz in th...

Page 2: ...1 Table of Contents Section Table of Contents Page 1 Safety Page 2 Contents Accessories Page 3 Specifications Page 4 Notes of Operations Page 5 Control Panel Setup Page 6 Operations Page 7 Warranty Pa...

Page 3: ...under the maximum speed when the density of the sample exceeds 1 2g cm the maximum speed of the rotor must be reduced accordingly When the centrifuge is running rotor is spinning Please keep an area o...

Page 4: ...ne tube cushions 6pc 4 Fuse 1pc 4 AMP 5 Operating manual 1pc Upon receipt please examine the package if there s anything wrong please kindly contact your dealer promptly 3 2 Accessories If you need op...

Page 5: ...ly Setting Time Range 1 60min Maximum RPM 4000 r min Maximum relative centrifugal force RCF 1000 rpm 112 RCF 1500 rpm 252 RCF 2000 rpm 447 RCF 2500 rpm 699 RCF 3000 rpm 1006 RCF 3500 rpm 1370 RCF 4000...

Page 6: ...en lid and remove specimen tubes Use only with accessories provided or parts designed specifically for use with this model Do not use centrifuge when the door is open Make sure the specimen tubes are...

Page 7: ...nate display between time speed On Off run stop 6 2 Setup 1 Open box and remove centrifuge from packaging 2 Place centrifuge onto flat surface 3 Open lid and remove foam packaging inside 4 Plug Power...

Page 8: ...etting for speed is in RPM s Please select desired speed shown in RPM s by using the or arrow buttons until the display panel reads your desired settings Spinning Centrifuge 2 Insert the centrifuge tu...

Page 9: ...air technicians or damage occurring in transit If you have questions concerning this product or warranty contact the dealer from whom it was purchased For warranty service product should be well packe...