background image

AeroPro

Cordless Prophy System

Directions for use

Please read carefully and completely  

before operating unit.

Summary of Contents for AeroPro

Page 1: ...AeroPro Cordless Prophy System Directions for use Please read carefully and completely before operating unit...

Page 2: ...3 Pre Operational Checks 9 4 4 Operating your AeroPro Handpiece 10 4 5 Changing Speed Modes 11 4 6 Charging the Battery ChargeSMART and Sleep Mode 11 4 7 Locking the Handpiece 13 4 8 Infection Prevent...

Page 3: ...ylaxis procedures This system was engineered with ChargeSMART technology to provide the clinician a full day battery life Premier s handpiece provides the flexibility of using Premier s 2pro prophy an...

Page 4: ...vice do not sterilize the motor component charging station AC adapter or DPA Disinfect the motor component charging station or AC adapter using only the tested and approved disinfectants listed in Sec...

Page 5: ...nd adheres to the skin or clothing immediately wash it off with clean water This may cause skin irritation According to IEC 60601 1 UL60601 1 this device must not be used in the presence of a flammabl...

Page 6: ...d the user s authority to operate the equipment This Class A digital apparatus complies with Canadian ICES 003 FCC ID 2AQ77 2018 AP IC 24334 AEROPRO Model AP 18 PDPC Product Marketing Name AeroPro Sys...

Page 7: ...and 1 Box of 500 Disposable Barriers 1 Bag of 48 Units 2pro Disposable Prophy Angles Short Soft 3 Extra Set of O rings and Tool for Maintenance 6 Small O rings for the Outer Sheath and 3 Large O rings...

Page 8: ...o the internal handpiece component motor component It also houses the bearings and acts as an interface between a DPA and the motor component The autoclavable outer sheath features an on board user co...

Page 9: ...lowing prophy angle to swivel Dimpled Silicone Grip Improves tactile feel while preventing slippage Charging Station with ChargeSMART Technology Induction Charging System Safe and convenient Preserves...

Page 10: ...structions described below 4 1 3 Charging for the first time Prior to first use fully charge the handpiece using the AeroPro AC adapter provided Failure to use the supplied AC adapter might cause malf...

Page 11: ...Procedure Section 4 8 4 3 2 Insert the motor component into the AeroPro custom fitted disposable barrier Ensure that the nose of the motor component is fully aligned to the opening of the disposable...

Page 12: ...ow 500 rpm 500 rpm Medium 1500 rpm High 2800 rpm 2800 rpm The 2 Speed Mode allows the operator to go from low to high speed with one button touch factory setting or if a greater range of speeds is pre...

Page 13: ...d ON OFF MODE button for 3 seconds Two top green indicator lights flash and beep two times and the handpiece is switched to 2 speed mode 4 5 4 Once the speed mode is selected press and release ON OFF...

Page 14: ...y charged all five indicator lights will display in a steady state one red one yellow and three green NOTE Allow 2 hours to fully charge the battery 4 6 6 Indicator lights represent available power us...

Page 15: ...will automatically turn off and the indicator lights will turn off 4 7 Locking the Handpiece In the event the AeroPro cordless handpiece needs to be transported it must first be set into the Locked po...

Page 16: ...sheath and the stand in the AeroPro Cordless Prophy System are not sterile upon receipt and must be steam autoclave sterilized prior to use in accordance with the following instructions Limitations o...

Page 17: ...eath Stand in a separate paper plastic steam sterilizing pouch If using a sterilizing cassette ensure that the sterilizer s maximum load is not exceeded Steam Autoclave Sterilization Place bagged Oute...

Page 18: ...e Containment and Transportation Handle with care Cleaning Using a CaviWipe or equivalent CDC intermediate level of disinfectant wipe wipe charging station motor component and AC adapter Discard used...

Page 19: ...nsions Handpiece with Outer Sheath L 159 mm W 27 mm Operating Environment Ambient temperature Relative Humidity Atmospheric Pressure Altitude 10 C 40 C 50 F to 104 F 45 95 non condensing 80 106 kPa 2...

Page 20: ...ear Month Date of Manufacturing This product meets UL safety standard requirements Electromagnetic interference from the device is under limits approved by the Federal Communications Commission For in...

Page 21: ...es of AeroPro Cordless Prophy System Auto turn off of the handpiece will occur for the following reasons 6 2 1 Overcurrent Protection Yellow light will flash and the handpiece will automatically turn...

Page 22: ...ollow standard prophylaxis procedures DPA Starts to Turn Around the Axis During a Procedure 1 Ensure the DPA is not damaged by removing the DPA and inspecting the base for cracks or other damage 2 O r...

Page 23: ...Too much current is going through the handpiece 2 Do not use handpiece for several minutes and allow the current to return to the specified limits Red Yellow Lights Flashed and Cordless Handpiece Aut...

Page 24: ...flash 6 times See Section 4 7 for locking the handpiece How Do I Unlock AeroPro Handpiece MUST be removed from the charging station Press and Hold ON OFF Mode button for 3 seconds The unit will beep 6...

Page 25: ...e according to local laws and regulations The handpiece motor component includes a battery pack lithium ion battery Do not dispose of in a fire or source of heat Doing so may cause rupture of the batt...

Page 26: ...ment guidance RF Emissions CISPR 11 Group 1 The AeroPro uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby ele...

Page 27: ...601 1 2 IEC 61000 4 5 1 kV line s to line s 1 kV line s to line s Mains power quality should be that of a typical commercial or hospital environment Voltage Dips Short Interruptions and Voltage Variat...

Page 28: ...T Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Product The product is intended for use in an electromagnetic environment in which radiated RF distur...

Page 29: ...be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE Theseguidelinesmaynotapplyinallsituations Electrom...

Page 30: ...le Outer Sheath 5500520 AeroPro Disposable Barriers Box of 500 5500530 AeroPro Charging Station 5500540 AeroPro AC Adapter 5500541 AeroPro Motor Component 5500550 AeroPro Stand 5500560 AeroPro O ring...

Page 31: ...y will void the foregoing warranty Parts will be repaired or replaced at the discretion of Premier provided that the system has been operated and maintained as prescribed in these instructions and has...

Page 32: ...novative standard setting solutions 4 Be your Product Concierge for knowledge and service 4 Make it right if we fall short Manufacturer Premier Dental Products Company 1710 Romano Drive Plymouth Meeti...

Reviews: