12
2. Wash canister, lid, o-ring, and overflow valve in a solution of warm water
with a mild, liquid detergent (e.g. Dawn or Palmolive) and rinse with clean,
warm tap water.
3. For single patient use: After washing, disinfect using one of the following
method:
Soak in 1 part vinegar (≥5% acetic acid concentration) to 3 parts water
(131°F-149°F/55°C-65°C) solution for 60 minutes. Rinse with clean, warm
water and air dry in a clean environment.
Suction Device (single-patient use)
1. With the power switch in the “Off” position, disconnect the PM66PED
EasyGo Vac from all external power sources.
2. Wipe the housing with a clean cloth and any commercial
(bacterial-germicidal) disinfectant.
CAUTION
Do not submerge in water as this will result in damage to the vacuum pump.
NOTE:
Do not use any cleaners or disinfectants that contain ammonia,
benzene and/or acetone to clean the device.
Suction Device (multi-patient use)
1. Dispose of all accessory components that are not suitable for reuse, i.e.,
collection canister, filter, tubing and carrying case.
2. With the power switch in the “Off” position, disconnect the PM66PED
EasyGo Vac from all external power sources.
3. Visually inspect device for any damage, missing parts, etc.
4. Wipe the housing with a clean cloth and a commercial (bacterial-germicidal)
disinfectant.
5. Replace all accessory components, i.e., collection canister, filter and tubing
with new.
CAUTION
Do not submerge in water as this will result in damage to the vacuum pump.
NOTE:
Do not use any cleaners or disinfectants that contain ammonia,
benzene and/or acetone to clean the device.