3

dANGER

• 

Caution must be exercised when using this Compressor where 

oxygen is being administered.

•  dO NOT

 use near any type of flame or flammable/explosive 

substances.

•  dO NOT

 allow the Compressor to exceed Operating or Storage 

Temperature specifications.

TO REDUCE THE RISK OF ELECTROCUTION

• 

Always unplug the Compressor immediately after using.

•  dO NOT 

use while bathing.

•  dO NOT 

place or store the Compressor where it can fall or be 

pulled into a tub or sink.

•  dO NOT 

place or drop into water or other liquid.

•  dO NOT 

reach for the Compressor that has fallen into water. 

Unplug immediately.

•  dO NOT

 attempt to repair the Easy Air Compressor. Any attempt 

could result in electrical shock.

WARNING

To reduce risk of burns, electrocution, fire, or injury to persons:

• 

Close supervision is necessary when this Compressor is used by, 

on, or near children or invalids.

• 

Use this Easy Air Compressor only for its "Intended Use" as 

described in this manual.

•  dO NOT

 use attachments not recommended by Precision Medical, Inc.

•  dO NOT

 use the Compressor with any equipment that requires "dry" 

air, including ventilators, gas blenders, or flow generators, without a 

device to properly dehumidify the discharged air of the PM15.

•  dO NOT 

use if the Compressor has a damaged cord or plug.

•  dO NOT

 use if the Compressor has been dropped or damaged.

• 

Keep the cord away from heated surfaces.

•  dO NOT

 use when sleepy or drowsy.

•  dO NOT

 drop or insert any object into any opening or hole.

•  dO NOT

 cover air vents. Restriction of air flow will cause unit to 

overheat, causing the thermal reset to shut down the Compressor.

Summary of Contents for EASY AIR 15

Page 1: ...n PM15EF PM15EP SAVE THESE INSTRUCTIONS CAUTION Federal USA law restricts this device to sale by or on the order of a physician 300 Held Drive Tel 001 610 262 6090 Northampton PA 18067 USA Fax 001 610...

Page 2: ...EFORE USING 2 Explanation of Abbreviations 2 Safety INFORMATION Warnings and cautions 2 SPECIFICATIONS 4 Component Identification 5 Grounding Instructions 6 OPERATING INSTRUCTIONS 6 Cleaning maintenan...

Page 3: ...nderstand this manual DO NOT USE the Easy Air Compressor and contact your Provider Explanation of Abbreviations psi Pounds Per Square Inch lpm Liters Per Minute Safety INFORMATION Warnings and caution...

Page 4: ...NING To reduce risk of burns electrocution fire or injury to persons Close supervision is necessary when this Compressor is used by on or near children or invalids Use this Easy Air Compressor only fo...

Page 5: ...40 C Storage Temperature Range 4 F to 104 F 20 C to 40 C Storage Operating Humidity Maximum 95 Noncondensing Safety Systems High Temperature Motor Shutdown Thermal Protector Class B Thermal Limit Aut...

Page 6: ...CAUTION Missing or illegible labels must be replaced contact Precision Medical Inc Carry Handle Air Inlet Filters Regulator Knob Gauge Power Cord On Off I O Switch Hour Meter Power Cord Keeper Air Ou...

Page 7: ...pressor near a properly grounded electrical receptacle Allow a minimum of 4 in 10 cm clearance on all sides to ensure adequate air flow for cooling of the Compressor 3 Be sure the On Off I O Switch is...

Page 8: ...or Exterior Components Use a clean cloth or sponge dampened with soap and water Wipe dry with a clean cloth or paper towel CAUTION Excessive dust buildup on filter will reduce performance of Compresso...

Page 9: ...Refer to Precision Medical Inc Return Policy available on the Internet www precisionmedical com NOTE When returning this Easy Air Compressor disconnect the power cord from the back of the unit disposa...

Page 10: ...Shut Compressor Off Wait 15 minutes Ensure vents are clear 4 Replace Fuse 5 Turn off Compressor When Power restored repeat Start Up Procedure 6 BEFORE STARTING UNIT Pull OUT regulator Knob to unlock...

Page 11: ...s about the merchandise described in this contract and any such statements shall not be relied upon and are not part of the contract for sale Thus this writing is a final complete and exclusive statem...

Page 12: ...edical Inc 300 Held Drive Northampton PA 18067 USA Toll Free Tel 800 272 7285 Toll Free Fax 800 353 1240 Tel 001 610 262 6090 Fax 001 610 262 6080 www precisionmedical com 502597 Rev12 10 08 10 M Prin...

Reviews: